background image

Valor™ 2000W Series

                                                                           

EN-9

                

 

4.4 Weigh Mode 

1. 

Press and hold 

Mode

 until [

wJeIGH

] (Weigh) is displayed. 

2. 

If required, place an empty container on the pan and press 

Tare

3. 

Add sample to the pan or container. The display shows the weight of the 

 sample. 

 

4.5 Checkweigh Mode  

This mode sets low and high weight limits for portion control processes.

 
1. Press and hold 

Mode

 until

 [

CHeCk

(Check) is displayed.

 [

CLr.ref

(clear 

references) will then display. 

 

Notes:

 Press 

Function

 to view the low and high reference weight limits. 

     

 If there is no reference data stored,

 [

Set.ref

is displayed. Press 

Yes

 to 

   

 set reference data. 

2.  Press 

No

 to use the stored reference weight limits and proceed to step 5.

 

3.  Press 

Yes

 to establish new reference values. The scale will then display

 [

Set. 

LO

]. 

Press 

Yes

 to view the “Low” limit value. Press 

Yes

 to accept or 

No

 to edit 

the “Low” limit value. The stored value then displays with the first digit 
highlighted

 [

0

00.000

 

kg

]. Repeatedly press 

No

 until the desired number 

appears. Press 

Yes

 to accept and highlight the next digit. Repeat until all the 

digits are correct. Press 

Yes

 to accept the “low” limit value,

 [

Set. HI

will be 

displayed. 

4.  Repeat the same procedure to accept or edit the “high” value. 
5. If required, place an empty container on the pan and press 

Tare

. Place 

sample material on the pan or in the container. If the sample weight is under 
the target weight range, the yellow LED will light. If the sample is within the 
target weight range, the green LED will light. If the sample is over the target 
weight range, the red LED will light. 

6. To clear the stored reference values press and hold 

Mode

 until

 [

CHeCk

] is 

displayed. Press 

Yes

 when

 [

Clr.ref

is displayed.

  

                 

 

 

 

 

 

 

 

 

 

    

www.

GlobalTestSupply

.com

Find Quality Products Online at:

sales@GlobalTestSupply.com

Summary of Contents for Valor 2000W Series

Page 1: ...ie Valor 2000W Manual de instrucciones S rie Valor 2000W Manuel d instructions Valor 2000W Serie Bedienungsanleitung Serie Valor 2000W Manuale di istruzioni www GlobalTestSupply com Find Quality Produ...

Page 2: ...conditions specified in these instructions Service should be performed only by authorized personnel Only use weights within the scale s capacity as specified in these instructions Do not operate the s...

Page 3: ...2 Level Indicator 2 5 Power The AC Adapter is used to power the scale when battery power is not needed First connect the AC Adapter plug to the scale input jack located at the bottom of the scale acco...

Page 4: ...battery operation the battery symbol indicates the battery charging status When charging the symbol will blink slowly and when fully charged the symbol will stop blinking TABLE 2 1 Symbol Charge Level...

Page 5: ...ting Short Press Enter clear a Tare value Extended Press3 Initiates Calibration procedure Short Press in Menu Toggles through available settings Notes 1 Short Press Press less than 2 5 seconds 2 Long...

Page 6: ...s displayed 3 3 Calibration Menu To enter the calibration menu extend press Cal button for 5 seconds SpaN will then be displayed Press Zero key to perform span calibration or press Cal key to proceed...

Page 7: ...ow instructions to enter the service menu to set the displayed unit With the scale off and LFT off extend press both the On Zero and Tare keys at the same time for more than 8 seconds until rAmMP is d...

Page 8: ...oggles through active Units Short Press in Menu Toggles through available settings Short Press Initiates an application mode specific response Long Press Selects active Mode Short Press in Menu return...

Page 9: ...nstalled it should be calibrated to ensure accurate results Before performing the calibration be sure to have the appropriate calibration weights as listed in table 3 3 Press and hold Menu until mMeNU...

Page 10: ...the Low limit value Press Yes to accept or No to edit the Low limit value The stored value then displays with the first digit highlighted 000 000 kg Repeatedly press No until the desired number appear...

Page 11: ...cates factory default setting Note The Calibration Menu Mode indicator is displayed when in the Menu Mode 4 6 2 Cal Menu Enter this menu to perform calibrations Span SpaN yes no Initiates a span calib...

Page 12: ...40 05 15 15 14 14 13 13 12 12 11 11 10 40 05 41 12 16 15 15 14 14 13 13 12 12 11 11 41 12 42 19 16 16 15 15 14 14 13 13 12 12 11 42 19 43 26 17 16 16 15 15 14 14 13 13 12 12 43 26 44 32 17 17 16 16 15...

Page 13: ...rtment 5 Remove the plastic pan 6 Clean the plastic pan 6 3 1 2 4 5 Figure 5 1 Pan cleaning 5 2 2 Reinstalling the pan after cleaning Please follow the steps below to reinstall the plastic pan see als...

Page 14: ...of range Tare value exceeds maximum Err 9 Data Internal data error Contact an authorized service agent Err13 mMEmM Fail to write EEPROM Contact an authorized service agent Err 53 C SUmM Invalid checks...

Page 15: ...lay 2 x Red LED front and rear 6 digit 7 segment 20 5 mm 0 8 in characters Keyboard V22PW 2 buttons V22XW 4 buttons Application Modes V22PW Weighing V22XW Weighing Checkweighing Battery Operating Time...

Page 16: ...irective 2006 66 EC introduces new requirements from September 2008 on removability of batteries from waste equipment in EU Member States To comply with this Directive this device has been designed fo...

Page 17: ...standard s requirements On June 21 2012 OHAUS Corporation USA was re registered to the ISO 9001 2008 standard Limited Warranty OHAUS products are warranted against defects in materials and workmanship...

Reviews: