background image

EN-10                                                                         Valor™ 2000W Series

 

 

4.6 Menu Settings  

The User Menu allows the customizing of scale settings.  

Note:

 Valor 2000W – V22PW does not have these Menus. 

 

4.6.1 Menu Navigation 

User Menu: 

Menu: 

C.A.L M.O.d.E U.n.i.t 

E.n.d 

 
 

Menu 

Items: 

Span 

Lin 

GEO 

End 

Reset 

Check 

End 

kg 

oz 

lb 

lb:oz 

End 

 

To Enter the Menu Mode 

Press and hold 

Menu

 until

 [

mMeNU

(Menu) is displayed. When released the first sub-

menu

 [

C.a.l

(Cal) will be shown.

  

Press 

Yes

 to enter the displayed sub-menu or press 

No

 to advance to the next. 

Selecting a sub-menu will display the first menu item. Press 

Yes

 to view the menu 

item setting or press 

No

 to move to the next menu item. When viewing the setting, 

press 

Yes

 to accept the setting, or press 

No

 to change the setting. When

 [

End

is 

displayed, press 

Yes

 to return to the sub-menu selections or 

No

 to return to the first 

item in the current menu. 

Bold

 indicates factory default setting. 

Note:

 

The Calibration / Menu Mode indicator is displayed when in the Menu Mode.

 

 

4.6.2 Cal Menu

 

Enter this menu to perform calibrations. 

 

Span

 

[

SpaN

(yes, no)  

 Initiates a span calibration procedure (zero and span). 

 

Lin

 

[

LIN

]

 (yes, no) 

 Initiates a linearity calibration procedure (zero, mid-point and span). 

 

GEO

 

[

GEO

]

  

 Geographical Adjustment Factor (GEO) is used to adjust the calibration    
 based on the current location. Settings from 0 to 31 are available with 12   
 being the default. Refer to table 4-2 to determine the GEO factor that   
 corresponds to your location. 

 

End Cal

 [

End

 Advance to the next menu or return to the top of the current menu.

 

 

4.6.3 Mode Menu  

Enter this menu to activate modes so they will be available for use with the Mode 
button. Weigh mode is always active by default. 

 

Reset

 

[

reset

]

 (

no

, yes) 

   Reset the Mode menu to factory defaults. 

 

Check

 

[

CHECk

]

 (off, 

on

   Set the sub-mode. 

 

End Mode

 

[

End

]

  

   Advance to the next menu or return to the top of the current menu. 

www.

GlobalTestSupply

.com

Find Quality Products Online at:

sales@GlobalTestSupply.com

Summary of Contents for Valor 2000W Series

Page 1: ...ie Valor 2000W Manual de instrucciones S rie Valor 2000W Manuel d instructions Valor 2000W Serie Bedienungsanleitung Serie Valor 2000W Manuale di istruzioni www GlobalTestSupply com Find Quality Produ...

Page 2: ...conditions specified in these instructions Service should be performed only by authorized personnel Only use weights within the scale s capacity as specified in these instructions Do not operate the s...

Page 3: ...2 Level Indicator 2 5 Power The AC Adapter is used to power the scale when battery power is not needed First connect the AC Adapter plug to the scale input jack located at the bottom of the scale acco...

Page 4: ...battery operation the battery symbol indicates the battery charging status When charging the symbol will blink slowly and when fully charged the symbol will stop blinking TABLE 2 1 Symbol Charge Level...

Page 5: ...ting Short Press Enter clear a Tare value Extended Press3 Initiates Calibration procedure Short Press in Menu Toggles through available settings Notes 1 Short Press Press less than 2 5 seconds 2 Long...

Page 6: ...s displayed 3 3 Calibration Menu To enter the calibration menu extend press Cal button for 5 seconds SpaN will then be displayed Press Zero key to perform span calibration or press Cal key to proceed...

Page 7: ...ow instructions to enter the service menu to set the displayed unit With the scale off and LFT off extend press both the On Zero and Tare keys at the same time for more than 8 seconds until rAmMP is d...

Page 8: ...oggles through active Units Short Press in Menu Toggles through available settings Short Press Initiates an application mode specific response Long Press Selects active Mode Short Press in Menu return...

Page 9: ...nstalled it should be calibrated to ensure accurate results Before performing the calibration be sure to have the appropriate calibration weights as listed in table 3 3 Press and hold Menu until mMeNU...

Page 10: ...the Low limit value Press Yes to accept or No to edit the Low limit value The stored value then displays with the first digit highlighted 000 000 kg Repeatedly press No until the desired number appear...

Page 11: ...cates factory default setting Note The Calibration Menu Mode indicator is displayed when in the Menu Mode 4 6 2 Cal Menu Enter this menu to perform calibrations Span SpaN yes no Initiates a span calib...

Page 12: ...40 05 15 15 14 14 13 13 12 12 11 11 10 40 05 41 12 16 15 15 14 14 13 13 12 12 11 11 41 12 42 19 16 16 15 15 14 14 13 13 12 12 11 42 19 43 26 17 16 16 15 15 14 14 13 13 12 12 43 26 44 32 17 17 16 16 15...

Page 13: ...rtment 5 Remove the plastic pan 6 Clean the plastic pan 6 3 1 2 4 5 Figure 5 1 Pan cleaning 5 2 2 Reinstalling the pan after cleaning Please follow the steps below to reinstall the plastic pan see als...

Page 14: ...of range Tare value exceeds maximum Err 9 Data Internal data error Contact an authorized service agent Err13 mMEmM Fail to write EEPROM Contact an authorized service agent Err 53 C SUmM Invalid checks...

Page 15: ...lay 2 x Red LED front and rear 6 digit 7 segment 20 5 mm 0 8 in characters Keyboard V22PW 2 buttons V22XW 4 buttons Application Modes V22PW Weighing V22XW Weighing Checkweighing Battery Operating Time...

Page 16: ...irective 2006 66 EC introduces new requirements from September 2008 on removability of batteries from waste equipment in EU Member States To comply with this Directive this device has been designed fo...

Page 17: ...standard s requirements On June 21 2012 OHAUS Corporation USA was re registered to the ISO 9001 2008 standard Limited Warranty OHAUS products are warranted against defects in materials and workmanship...

Reviews: