EN-6                                                                           Valor™ 2000W Series

                

 

3.4 Initial Calibration

 

When the scale is first installed it should be calibrated to ensure accurate results. 
Before performing the calibration, be sure to have the appropriate calibration 
weights as listed in table 3-3. 
Press and hold 

Cal 

for 5 seconds, the display will show [

Cal

]. When the button is 

released, the display will show

 [

SpaN

]. Press 

On/Zero

 to begin the span calibration. 

[

--C--

blinks while zero reading is stored. Next, the display shows the calibration 

weight value. Place the specified calibration mass on the pan and press 

On/Zero

.

 

[

--C--

blinks while the reading is stored. The scale returns to the previous 

application mode and is ready for use.  
The message [

Err 3.0  Cal

] will be displayed if these calibration steps are not 

followed or if the wrong weight was used. 
The calibration process can be aborted by turning the scale off.

TABLE 3-3 

Required Span Calibration Mass 

(sold separately)

 

Max 

Mass

Max 

Mass

1500 g 

1 kg / 3 lb 

15000 g 

10 kg / 30 lb

3000 g 

2 kg / 6 lb 

30000 g 

20 kg / 60 lb

6000 g 

5 kg / 15 lb 

 

 

            

Note:

 If current unit is lb or oz, the scale will be calibrated by pound weight. 

3.5 Unit Menu

 

Follow below instructions to enter the service menu to set the displayed unit. 
With the scale off and LFT off, extend press both the 

On/Zero

 and 

Tare

 keys at the 

same time for more than 8 seconds until [

rAmMP

] is displayed. Press the 

Tare

 key 

through the succeeding displays: [

rAmMP

]*, [

LIN

]*, [

SPAN

]*, [

CAP

]*, [

Dius

]* until [

s.UNIt

is displayed. 

 When 

[

s.UNIt

] is displayed, press the 

On/Zero

 key.   

 

[

UNIt 

kg

] will be displayed.  Press the 

Tare

 key until the active (power on) 

unit is displayed. 

 

When the active unit is displayed, press the 

On/Zero

 key.  

 

[

ON 

XX

] is displayed.  Press 

Tare

 to switch to [

Off 

XX

], then press the 

On/Zero

 key**. 

 

The next unit [

UNIt 

xx

] is displayed.  Press the 

Tare

 key until the desired 

unit is displayed. 

 

When the desired unit is displayed, press the 

On/Zero

 key. 

 

[

Off 

XX

] is displayed.  Press the 

Tare

 key to switch to [

ON 

XX

], then press the 

On/Zero

 key. 

 Press 

the 

Tare

 key until [

End

] is displayed, then press the 

On/Zero

 key. 

 Press 

the 

Tare

 key through the succeeding displays: [

e.paND

]*, [

reset

]* 

until [

End

] is displayed. 

When [

End

] is displayed, press the 

On/Zero

 key to exit to weighing mode.  

Notes:  

*Do not change any of these settings (by pressing the On/Zero key). This may affect 
the correct set up of the scale.  
**The active unit must be turned off first before another unit can be turned on. 

www.

GlobalTestSupply

.com

Find Quality Products Online at:

sales@GlobalTestSupply.com

Summary of Contents for Valor 2000W Series

Page 1: ...ie Valor 2000W Manual de instrucciones S rie Valor 2000W Manuel d instructions Valor 2000W Serie Bedienungsanleitung Serie Valor 2000W Manuale di istruzioni www GlobalTestSupply com Find Quality Produ...

Page 2: ...conditions specified in these instructions Service should be performed only by authorized personnel Only use weights within the scale s capacity as specified in these instructions Do not operate the s...

Page 3: ...2 Level Indicator 2 5 Power The AC Adapter is used to power the scale when battery power is not needed First connect the AC Adapter plug to the scale input jack located at the bottom of the scale acco...

Page 4: ...battery operation the battery symbol indicates the battery charging status When charging the symbol will blink slowly and when fully charged the symbol will stop blinking TABLE 2 1 Symbol Charge Level...

Page 5: ...ting Short Press Enter clear a Tare value Extended Press3 Initiates Calibration procedure Short Press in Menu Toggles through available settings Notes 1 Short Press Press less than 2 5 seconds 2 Long...

Page 6: ...s displayed 3 3 Calibration Menu To enter the calibration menu extend press Cal button for 5 seconds SpaN will then be displayed Press Zero key to perform span calibration or press Cal key to proceed...

Page 7: ...ow instructions to enter the service menu to set the displayed unit With the scale off and LFT off extend press both the On Zero and Tare keys at the same time for more than 8 seconds until rAmMP is d...

Page 8: ...oggles through active Units Short Press in Menu Toggles through available settings Short Press Initiates an application mode specific response Long Press Selects active Mode Short Press in Menu return...

Page 9: ...nstalled it should be calibrated to ensure accurate results Before performing the calibration be sure to have the appropriate calibration weights as listed in table 3 3 Press and hold Menu until mMeNU...

Page 10: ...the Low limit value Press Yes to accept or No to edit the Low limit value The stored value then displays with the first digit highlighted 000 000 kg Repeatedly press No until the desired number appear...

Page 11: ...cates factory default setting Note The Calibration Menu Mode indicator is displayed when in the Menu Mode 4 6 2 Cal Menu Enter this menu to perform calibrations Span SpaN yes no Initiates a span calib...

Page 12: ...40 05 15 15 14 14 13 13 12 12 11 11 10 40 05 41 12 16 15 15 14 14 13 13 12 12 11 11 41 12 42 19 16 16 15 15 14 14 13 13 12 12 11 42 19 43 26 17 16 16 15 15 14 14 13 13 12 12 43 26 44 32 17 17 16 16 15...

Page 13: ...rtment 5 Remove the plastic pan 6 Clean the plastic pan 6 3 1 2 4 5 Figure 5 1 Pan cleaning 5 2 2 Reinstalling the pan after cleaning Please follow the steps below to reinstall the plastic pan see als...

Page 14: ...of range Tare value exceeds maximum Err 9 Data Internal data error Contact an authorized service agent Err13 mMEmM Fail to write EEPROM Contact an authorized service agent Err 53 C SUmM Invalid checks...

Page 15: ...lay 2 x Red LED front and rear 6 digit 7 segment 20 5 mm 0 8 in characters Keyboard V22PW 2 buttons V22XW 4 buttons Application Modes V22PW Weighing V22XW Weighing Checkweighing Battery Operating Time...

Page 16: ...irective 2006 66 EC introduces new requirements from September 2008 on removability of batteries from waste equipment in EU Member States To comply with this Directive this device has been designed fo...

Page 17: ...standard s requirements On June 21 2012 OHAUS Corporation USA was re registered to the ISO 9001 2008 standard Limited Warranty OHAUS products are warranted against defects in materials and workmanship...

Reviews: