CHAPTER 1 GETTING STARTED
Ohaus Corporation www.ohaus.com
1-6
Valor
™ 2000/4000 Series Service Manual
1.7 SCALE OPERATION
– VALOR 2000
This section contains information on the basic operation of the Valor 2000 scale.
1.7.1 OVERVIEW OF THE CONTROLS
Figure 1-3. Valor 2000 Front Display with controls.
TABLE
1-6. Control functions Valor 2000
Button
Functions
Short Press
1
(when on): Sets display to zero
Long Press
2
(when off): Turns scale on
(when on): Turns the scale off
Short Press (in Menu): Selects / accepts displayed setting
Short Press: Enter / clear a Tare value
Long Press: Initiates Calibration procedure
Short Press (in Menu): Toggles through available settings
Extended Press
3
: Enters user menu
Notes:
1
Short Press: Press less than 2.5 seconds.
2
Long Press: Press and hold for more than 2.5 seconds.
3
Extended Press: Press and hold for more than 5 seconds.
On/Zero
Off
Tare
Cal
Summary of Contents for VALOR 4000
Page 3: ......