CHAPTER 2 DIAGNOSTIC GUIDE
Valor
™ 2000/4000 Series Service Manual
2-3
Ohaus Corporation www.ohaus.com
CAUTION:
IN THE NEXT STEP, DO NOT OVERLOAD THE SCALE BEYOND FULL
CAPACITY RATING.
– Measure the voltages on +SIG and –SIG wires, disconnected from PCB.
Note:
Measurements must be made with these wires disconnected from the
PCB. These measurements represent the output of the Load Cell. Record
measurements at Zero Load, 50% and full scale capacities. See Table 2-3 for
typical readings.
Figure 2-1. Top view of main PCB.
NOTE
: Table 2-3 indicates typical readings. Actual values can vary, but should
remain linear throughout the range. If readings are out of tolerance, replace the
Load cell. (See Section 3.6.)
TABLE 2-3. LOAD CELL OUTPUT READINGS (in mV/V with 4.65 V Excitation)
Model/Capacity
Zero Load
50% Load
100% Load
V41xWE1501T / 1.5kg
V22xWE1501T / 1.5kg
0.739
1.489
2.239
V41xWE3T / 3kg
V22xWE3T / 3kg
0.4434
1.3434
2.2434
V41xWE6T / 6kg
V22xWE6T / 6kg
0.2217
1.1217
2.0217
V41xWE15T / 15kg
V22xWE15T / 15kg
0.11085
1.23585
2.36085