CHAPTER 3 MAINTENANCE PROCEDURES
Ohaus Corporation www.ohaus.com
3-6
Valor
™ 2000/4000 Series Service Manual
TABLE 3-2. VALOR 2000/4000 OVERLOAD STOP GAP SETTINGS
Model
Max. Capacity
Overload Stop Gap
Scale
Load Cell
A/B mm
C/D mm
V41xWE1501T
V22xWE1501T
1.5kg
3kg
1.1
0.9
V41xWE3T
V22xWE3T
3kg
5kg
1.2
0.8
V41xWE6T
V22xWE6T
6kg
10kg
1
0.7
V41xWE15T
V22xWE15T
15kg
20kg
1
0.7
Figure 3-6. Gap position of corners A, B, C, D. (A & B represent the front of the scale.)