CHAPTER 3 MAINTENANCE PROCEDURES
Valor
™ 2000/4000 Series Service Manual
3-7
Ohaus Corporation www.ohaus.com
3.7 Removing/Replacing the Rechargeable Battery
The Valor 2000/4000 has a
rechargeable battery.
If it fails to recharge, replace it as
follows:
1.
Lift the battery’s front end
enough to access the quick-
connect tabs holding the wire
leads.
Note:
Be careful not to short-
circuit the battery leads.
2. Pull the quick-connect tabs
free of the battery posts,
and lift out the battery.
3. Position the replacement battery in the compartment on an angle sufficient to
access the battery posts. Make sure that the red harness connect to the red post;
and black harness connect to the black post.
4. Insert the quick-connect tabs on their respective posts, and place the battery fully in
the battery compartment.
3.8 Replacing the Function Label
The Function Label may need to be replaced. (See Chapter 5 for parts information.)
Use a broad knife to remove the label. Clean the glue residue from the Housing surface.
Then carefully place the new label where the old one was.