APPENDIX B SERVICE CALIBRATION
Ohaus Corporation www.ohaus.com
B-4
Valor
™ 2000/4000 Series Service Manual
B.7 DivS
When
DIVs
appears, press
YES
to accept the DivS menu item,
or press
NO
to advance to the next menu item.
Set the division to either HI or Low.
HIGH = 6000 / 7500 resolution
LOW = 3000 resolution
DIVs
Note: This is only for non-LFT mode. For LFT mode the resolution will
be set to 3000 e automatically.
B.8 S.Unit
When
Unit
appears, press
YES
to accept the Unit menu item,
or press
NO
to advance to the next menu item.
The available units are: kg, g, lb, oz, lb:oz.
Press YES to set the displayed unit or NO to advance to the next unit.
Each unit can be set to either ON or OFF.
ON = enabled
OFF = disabled
Note:
For Valor 2000, when changing unit, first set the current unit to
OFF then set the desired unit to ON.
S.Unit
B.9 Expand Menu
When
e.paNd
appears, press
YES
to accept the Expand menu, or
No
to advance to the Reset menu item.
OFF = disabled
ON = enabled
Note:
Expand is required for various performance tests.
e.paNd
B.10 Reset Menu
When
reset
appears, press
YES
to accept the Reset menu, or
No
to
advance to the End menu item
YES = Reset the Service menu to factory defaults.
NO = keep current setting
reset
B.11 End Menu
When
e.N.D
appears, press
YES
to return to exit the service menu, or
No
to advance to the CAP menu.
e.N.D