EN-13
Valor 5000 Series
3.7 Scale Settings
Enter the Menu mode. Proceed to a menu item and use the
Yes
and
No
buttons to view and select
available scale settings. When the desired changes have been made, press
Exit
to return to weighing.
Refer to Section 3.4 for more information on how to enter and navigate the menus.
3.7.1
Calibration sub-menu
Note
: Calibration is not available when the scale has been set to Legal for Trade operation (LFT set to
ON in the Setup sub-menu, refer to Section 3.7.2).
Calibration Types – Calibration can be adjusted in two ways, Span and Linearity. Span calibration
uses two calibration points to adjust the sensitivity of the scale. Linearity calibration uses three
calibration points to correct for non-linear weighing results.
Calibration Weights – Before beginning calibration, make sure that the required calibration weights are
available. The model-specific calibration points are listed in the table below.
TABLE 3-3. CALIBRATION POINTS.
Notes
:
1. Default calibration points are shown in bold print.
2. Calibration Units kg or lb are selected in the Setup sub-menu (Section 3.7.2.)
3. ASTM Class 4 or OIML Class F2 weights are required for calibration.
Model Calibration Unit Span Cal. points Linearity cal. points
V51P3 / V51PH3
kg 1, 2 or
3
kg
1
or 2 and
3
kg
lb 2, 5 or
6
lb
2
or 4 and
6
lb
V51P6 / V51PH6
kg 2, 5 or
6
kg
2
or 4 and
6
kg
lb 5, 10 or
15
lb
5
or 10 and
15
lb
V51P15 / V51PH15 kg 5, 10 or
15
kg
5
or 10 and
15
kg
lb 10, 20 or
30
lb
10
or 20 and
30
lb
V51P30 / V51PH30 kg 10, 20 or
30
kg
10
or 20 and
30
kg
lb 10, 30 or
60
lb
20
or 30 and
60
lb