CHAPTER 3 MAINTENANCE PROCEDURES
Ohaus Corporation www.ohaus.com
3-6
Valor
™
7000 Series Service Manual
TABLE 3-2. VALOR 7000 OVERLOAD STOP GAP SETTINGS
Model
Max. Capacity
Overload Stop Gap
V7000 Scale
Load Cell
A/B mm
C/D mm
V71P1501T
1.5kg
3kg
0.6
0.5
V71P3T
3kg
5kg
0.8
0.7
V71P6T
6kg
10kg
1.5
1.1
V71P15T
15kg
20kg
2.4
1.7
V71P30T
30kg
40kg
2.4
1.7
Figure 3-6. Gap position of corners A, B, C, D. (A & B represent the front of the scale.)
A
B
D
C
Load Cell