CHAPTER 4 TESTING
Valor
™
7000 Series Service Manual
4-1
Ohaus Corporation www.ohaus.com
4.1 TESTING
Before and after servicing a Valor 7000 scale, an operational test and various performance tests
should be made to confirm that the scale meets specifications. Turn the scale on and allow it to
warm up for at least one hour before performing these tests.
NOTE:
Make sure the test area is free from drafts and that the scale rests on a level and
vibration-free surface.
4.1.1 TEST MASSES REQUIRED
The masses required to test the Ohaus Valor 7000 scales must meet the requirements of ASTM
Class 4 or OIML F2 Tolerance. The mass values are listed in Table 4-1.
TABLE 4-1. TEST MASS VALUES
Model
Weight (g)
V71P1502T
1kg,500g,200g
V71P3T
1kg,500g
V71P6T
2kg,1kg,500g
V71P15T
10kg,5kg,2kg
V71P30T
10kg,5kg
4.2
Operational Test
1. Connect a functioning Power cord to the back of the scale.
2. Plug the Power Cord into a suitable power source, or power the scale on using
battery power. (Assure that the battery is charged beforehand.)
4.3
Segment Display Test
Turn the scale on, and ensure that all segments are enabled and displayed briefly and
followed by a software revision number.