APPENDIX B SERVICE CALIBRATION
Valor
™
7000 Series Service Manual
B-1
Ohaus Corporation www.ohaus.com
APPENDIX B.
SERVICE CALIBRATION
This section describes the Service Menu and sub-menus, which allow authorized service
personnel to perform factory Linearity and Span calibrations (no pre-set limits).
Note:
Be careful not to touch the scale or the surface it rests on while calibration is in
progress, as it will cause the process to fail.
B.1 Service Menu Structure
B.2 Entering the Service Menu
Turn the scale off.
Press the
On/Zero
then hold the
Function
and
M+
keys
simultaneously for 6 seconds until
S.E.r.V.C.e
is displayed.
S.E.r.V.C.e
Service
Cap
Span
Lin
Ramp
Expand
End
Grad