APPENDIX B SERVICE CALIBRATION
Ohaus Corporation www.ohaus.com
B-2
Valor
™
7000 Series Service Manual
B.3 Capacity Menu
When
CAP
appears, press
YES
to accept the Capacity menu item,
or press
NO
to advance to the menu item.
CAP
The current Capacity Setting is displayed, blinking.
(Outlined characters represent a flashing display.)
Note: The default setting is 3
Press
YES
to accept the setting and move to the Service Grad menu
item.
Press
NO
to enter another value using the numeric keypad and then
press
YES
to store the new value.
B.4 Graduation Menu
When
GrAd
appears, press
YES
to accept the Graduation menu or
No
to advance to the Service Span Calibration menu.
Set the graduation of the scale.
0.00005,…,0.02
Grad
B.5 Service Span Calibration
When
SPAN
appears, press
YES
to accept the Service Span
Calibration menu or
No
to advance to the Service Linearity
Calibration menu.
0
flashes. With no weight on the pan, press
Yes
to establish the
zero point. (Outlined characters represent a flashing display.)
- - C - -
flashes while the zero point is established.
Service Span Calibration point flashes.
(The example shows the value for a Scale Capacity of 3 kg.)
Place the specified calibration weight on the pan and press
Yes
.
- - C - -
flashes while the span point is established.
If span calibration was successful, the actual weight reading
appears for three seconds, followed by the Service Linearity
Calibration menu item.
(The example shows a 3 kg span weight on the pan.)
SPAN
--C--
—
--C--
—
3.000