APPENDIX B SERVICE CALIBRATION
Ohaus Corporation www.ohaus.com
B-4
Valor
™
7000 Series Service Manual
B.7 Ramp Menu
When
ramMp
appears, press
YES
to accept the Ramp menu, or
No
to advance to the Expand menu item.
The Ramp value is displayed as a percent of the A/D range.
Press
YES
to advance to the Expand menu item.
Note: Ramp is used to troubleshoot load cell problems. 0%
represents no signal from the A/D, 100% reflects the upper signal
range of the A/D.
ramMp
0.0
to
100.0
B.8 Expand Menu
When
e.paNd
appears, press
YES
to accept the Expand menu, or
No
to advance to the End menu item.
OFF = disabled
ON = enabled
Note: Expand is required for various performance tests.
e.paNd
B.9 End Menu
When
e.N.D
appears, press
YES
to return to exit the service menu,
or
No
to advance to the CAP menu.
e.N.D