background image

CHAPTER 1  GETTING STARTED 

Valor

 7000 Series Service Manual

  

1-5

 

Ohaus Corporation  www.ohaus.com

 

1.5.3  Menu Setup 

Programmable features of the Valor 7000 scales are contained in menus which are 

accessed  through  the  Display  Panel’s  control  switches.    See  the  I

nstruction Manual for a full 

description of the menus and how to access them. 

TABLE 1-4.  Valor 7000 MENU STRUCTURE 

Menu: 

C.A.L 

S.E.t.U.P 

r.E.A.d 

M.O.d.E 

 

 
 
Menu 
Items: 

Span 

Lin 

GEO 

End 

Reset 

Pwr.Un 

Zero 

A. Tare 

Bp.Sig 

Bp.Key 

IR Func 

IR Adj 

Accum 

End 

Reset 

Stable 

Filter 

AZT 

Light 

A.Off 

End 

Reset 

Weigh 

Percnt 

Check 

End 

 

 

 

 

U.n.i.t 

P.r.i.n.t 

C.O.M 

L.O.C.k 

E.n.d 

 
 
Menu 
Items: 

Reset 

kg 

lb 

oz 

lb:oz 

End 

Reset 

Stable 

A.Print 

Contnt 

Layout 

Data.Tr 

End 

Reset 

Baud 

Parity 

Stop 

Handsh 

Alt.Cm 

End 

L.Cal 

L.Setup 

L.Read 

L.Mode 

L.Unit 

L.Print 

L.COM 

End 

 

1.5.4  Menu Navigation  

Press and hold Menu until [

mMeNU

] (Menu) is displayed. When released the first sub-menu 

[

C.a.l

] (Cal) will be shown.  

Summary of button navigation functions in menu mode: 
 

-- 

Yes

    Allows entry into the displayed menu. 

 

 

 

   - Accepts the displayed setting and advances to the next item. 

 

-- 

No

   

Skips by the displayed menu. 

 

 

 

   - Rejects the displayed setting or menu item and advances to the next 

 

 

 

   available item.  

 

-- 

Back

  Moves backwards through the upper and middle level menus. 

 

 

 

   - Backs out of a list of selectable items to the previous middle level menu. 

     

-- 

Exit

 

Exits from menu directly to the active weighing mode.

 

 

 

 

 

 

Summary of Contents for Valor 7000 series

Page 1: ...SERVICE MANUAL VALOR 7000 Ohaus Corporation 7 Campus Drive Suite 310 Parsippany NJ 07054 973 377 9000...

Page 2: ...publication but Ohaus Corporation assumes no liability arising from the use or misuse of this material Reproduction of this material is strictly prohibited Material in this manual is subject to chang...

Page 3: ...3 Testing the Printed Circuit Board PCB 2 4 2 2 4 Diagnostic Guide 2 6 CHAPTER 3 MAINTENANCE PROCEDURES 3 1 Preventive Maintenance 3 1 3 1 1 Preventive Maintenance Checklist 3 1 3 2 Service Strategy...

Page 4: ...ucture 1 5 2 1 Load Cell Resistance Readings in Ohms 2 2 2 2 Color Code for Load Cell Wiring 2 2 2 3 Load Cell Output Readings in mV with 3 3V Excitation 2 3 2 4 Diagnostic Guide 2 6 3 1 Mounting Bolt...

Page 5: ...ponents Appendix A Standard Calibration Explains procedures for Standard Calibration performed prior to using a scale and after service Appendix B Service Calibration Describes the Service Menu and su...

Page 6: ...0 1 30000 LFT ON Resolution nMax 1 3000 1 3000 1 3000 1 3000 1 3000 Linearity 2d r0 0001 kg r0 0002 kg r0 0005 kg r0 001 kg r0 002 kg Repeatability 2d r0 0001 kg r0 0002 kg r0 0005 kg r0 001 kg r0 002...

Page 7: ...g Press Toggles through active weighing units Short Press in Menu Toggles through available settings Function Mode Back Short Press Selects function setting Long Press Selects active Mode Short Press...

Page 8: ...item 11 in figure 3 2 will remain lit The scale can be operated during charging and the battery is protected against overcharging For maximum operating time the battery should be charged at room tempe...

Page 9: ...O M L O C k E n d Menu Items Reset kg g lb oz lb oz End Reset Stable A Print Contnt Layout Data Tr End Reset Baud Parity Stop Handsh Alt Cm End L Cal L Setup L Read L Mode L Unit L Print L COM End 1 5...

Page 10: ...is not accessible Zero Range is locked at 2 Stable Range setting is locked at 1d Auto Zero Tracking setting is locked at 0 5d Units are locked at their current settings Filter is locked at current se...

Page 11: ...CHAPTER 1 GETTING STARTED Valor 7000 Series Service Manual 1 7 Ohaus Corporation www ohaus com Figure 1 3 Sealing...

Page 12: ...ith the most common cause first If the first remedy does not fix the problem proceed to the next remedy Before attempting to repair the balance read all chapters of this manual to be familiar with the...

Page 13: ...d full load output The reading should meet the Load Cell specifications The Load Cell output should be very close to linear over its capacity range NOTE The following steps involve power applied to th...

Page 14: ...ts at Zero Load 50 and full scale capacities See Table 2 3 for typical readings Figure 2 1 Top view of main PCB NOTE Table 2 3 indicates typical readings Actual values can vary but should remain linea...

Page 15: ...n 3 Using a DVM measure the excitation voltage EXE and EXE in Figure 2 1 The reading should be 3 1 volts dc This is the excitation voltage for the Load Cell and is regulated If the voltage is lower re...

Page 16: ...ue 6 Adjust the Simulator to simulate 0 load 50 load and 100 load for the capacity that the scale is rated for See Table 2 3 for values to use If the resulting readings are unstable the Main PC Board...

Page 17: ...below Power On Zero limit Err 8 3 Over Range Error Weight reading exceeds Overload limit Err 8 4 Under Range Error Weight reading below Underload limit Err 8 5 Tare out of range Adjust tare value to b...

Page 18: ...tronics to stabilize 3 1 1 Preventive Maintenance Checklist The scale should be inspected and checked regularly as follows 1 Remove the Pan and Sub Pan to inspect and clean the area beneath the Pan 2...

Page 19: ...the eight screws holding the Housing in place See Figure 3 1 3 Move the top housing a little bit backward and then raise and remove the top housing 4 Separate Top Housing from Bottom Housing Avoid str...

Page 20: ...PCB to the bottom housing 3 If either the Main PCB or the Load Cell is to be replaced note the order of the wire colors and then disconnect the Load Cell Cable from the Main PCB See Figure 3 2 4 Posit...

Page 21: ...washers that hold the Load Cell assembly frame and load cell to the Bottom Housing Remove the bolts and their washer pairs that hold the Load Cell Frame to the Load Cell Use a high leverage Allen wren...

Page 22: ...ual to the scale s capacity See Table 3 2 There are four Overload Stop Bolts See Figure 3 5 Adjust the Overload Stops per Table 3 2 1 Adjust the Overload Stop Bolt so that the gap between the Load Cel...

Page 23: ...VERLOAD STOP GAP SETTINGS Model Max Capacity Overload Stop Gap V7000 Scale Load Cell A B mm C D mm V71P1501T 1 5kg 3kg 0 6 0 5 V71P3T 3kg 5kg 0 8 0 7 V71P6T 6kg 10kg 1 5 1 1 V71P15T 15kg 20kg 2 4 1 7...

Page 24: ...2 Pull the quick connect tabs free of the battery posts and lift out the battery Figure 3 7 Rechargeable battery 3 Position the replacement battery in the compartment on an angle sufficient to access...

Page 25: ...S REQUIRED The masses required to test the Ohaus Valor 7000 scales must meet the requirements of ASTM Class 4 or OIML F2 Tolerance The mass values are listed in Table 4 1 TABLE 4 1 TEST MASS VALUES Mo...

Page 26: ...ication for each model listed in Table 1 1 1 Power on the balance The reading on the display should be 0g 2 Select a mass weighing near the maximum capacity of the balance and place it on the center o...

Page 27: ...test set up the instrument as follows Set Up Follow the steps in Appendix A Section A 2 Setup and Calibration Record Settings Zero Tracking Setting ____________ Displayed Units ____________ Mass Used...

Page 28: ...ge 6 Mean Reading 1 Reading 2 Reading 3 Reading 4 Reading 5 7 Reading 6 Reading 7 Reading 8 Reading 9 Reading 10 10 Mean ________ 6 Calculate the Delta for each reading and record in the work sheet De...

Page 29: ...press ON Zero Off 3 Place the test mass on the Scale record the weight and remove 4 Place Load 2 on the Scale and press ON Zero Off 5 Place the test mass on the Scale record the weight and remove 6 P...

Page 30: ...test weight in the center of the Weighing Pan 2 Tare the balance 3 Move the weight to point A and record the reading 4 Move the weight to point B and record the reading 5 Move the weight to point C an...

Page 31: ...d NOTE In all cases where a part is replaced the scale must be thoroughly checked after the replacement is made The scale MUST meet the parameters of all applicable specifications in this manual If fu...

Page 32: ...CHAPTER 5 PARTS LISTS DIAGRAMS Ohaus Corporation www ohaus com 5 2 Valor 7000 Series Service Manual 5 1 Valor 7000 SCALES PARTS Figure 5 1 Valor 7000 Scales Parts...

Page 33: ...der Down R31 R21 RC31 RC21 V71 8 30037420 Hardware Kit R31 R21 RC31 RC21 V71 9 30037400 Housing Bottom V71 10 30037421 Feet R31 R21 RC31 RC21 V71 11 30037422 Cover Option R31 R21 RC31 RC21 V71 12 3003...

Page 34: ...libration process 2 Allow the scale to warm up for approximately five minutes after stabilizing to room temperature 3 To abort calibration press Exit key or turn the scale off anytime during the calib...

Page 35: ...accept SpaN will then be shown Press Yes to begin the span calibration 0 kg will be displayed SpaN 0 3 Press Yes to accept C will be displayed while zero reading is stored Next the display shows the c...

Page 36: ...factory Linearity and Span calibrations no pre set limits Note Be careful not to touch the scale or the surface it rests on while calibration is in progress as it will cause the process to fail B 1 S...

Page 37: ...e Service Span Calibration menu Set the graduation of the scale 0 00005 0 02 Grad B 5 Service Span Calibration When SPAN appears press YES to accept the Service Span Calibration menu or No to advance...

Page 38: ...to indicate the calibration unit The example shows the mid calibration point value for a Scale Capacity of 3 kg Place the specified calibration weight on the pan and press Yes C flashes while the mid...

Page 39: ...o the Expand menu item Note Ramp is used to troubleshoot load cell problems 0 represents no signal from the A D 100 reflects the upper signal range of the A D ramMp 0 0 to 100 0 B 8 Expand Menu When e...