58
ENTER LIQUID DENSITY IN CC
Enter density of test liquid
Enter the density value for the test liquid. The standard density for water has
been pre-entered. Enter the correct value for the liquid you are using, and
press
Enter
button. Display advances to RETURN TO SETUP.
0.9971
0.9971
Return to setup
Press the
Enter
button. Display advances to PIPETTE SETUP with BARO
PRESSURE highlighted.
SELECT TEST LIQUID
H2O
OTHER
RETURN TO SETUP
Enter
Enter
PIPETTE SETUP
LIBRARY NAME
P2
TEST LIQUID(G/CC)
0.9971
BARO PRESSURE
1.0ATM
AUTO SAMPLE
OFF
AUTO PRINT
OFF
NUMBER OF SAMPLES
10
PIPETTE NAME
Enter barometric pressure
Press the
Enter
button. The display advances to BAROMETRIC PRESSURE
with ATM highlighted.
Select unit of atmospheric pressure measurement
Select unit, press
Enter
button. The display advances to BAROMETRIC
PRESSURE.
BAROMETRIC PRES-
SURE
ATM
PSIG
MMHG
MBAR
HPA
INHG
Select auto sampling
If you want auto sampling, press
Enter
button. Display advances AUTO
SAMPLE with ON highlighted. Press the
Enter
button.
Enter
BAROMETRIC PRESSURE
Specify barometric pressure
Enter the barometric pressure of the test area, then press the
Enter
button.
Display advances to PIPETTE SETUP with AUTO SAMPLE OFF highlighted. If
you do not want auto sampling, skip next two steps and select AUTO PRINT.
1.0000
1.000
It should be noted that in the next step you can select AUTO SAMPLE either
ON or OFF. When AUTO SAMPLE is turned ON, it is a real time saver as the
balance controls do not have to be touched during sampling.
PIPETTE SETUP
LIBRARY NAME
P2
TEST LIQUID(G/CC)
0.9971
BARO PRESSURE
1.0ATM
AUTO SAMPLE
OFF
AUTO PRINT
OFF
NUMBER OF SAMPLES
10
PIPETTE NAME
Select auto print on
Press the
Enter
button. The display advances to PIPETTE SETUP with AUTO
PRINT OFF highlighted.
Enter
PIPETTE SETUP
LIBRARY NAME
P2
TEST LIQUID(G/CC)
0.9971
BARO PRESSURE
1.0ATM
AUTO SAMPLE
OFF
AUTO PRINT
OFF
NUMBER OF SAMPLES
10
PIPETTE NAME
Balance modes