59
Set auto print on
When set ON, results are automatically printed. Press the
Enter
button.
Display advances to PIPETTE SETUP with NUMBER OF SAMPLES highlighted.
Enter
Enter
PIPETTE SETUP
LIBRARY NAME
P2
TEST LIQUID(G/CC)
0.9971
BARO PRESSURE
1.0ATM
AUTO SAMPLE
OFF
AUTO PRINT
OFF
NUMBER OF SAMPLES
10
PIPETTE NAME
Select number of samples
Press the
Enter
button. The display advances to NUMBER OF SAMPLES.
Enter sample size
Enter the number of samples. Press
Enter
button. The display advances to
PIPETTE SETUP with PIPETTE NAME highlighted.
AUTO PRINT
ON
OFF
10
1 10
NUMBER OF SAMPLES
Select pipette name
Press
Enter
button. Display advances to new screen PIPETTE NAME.
Enter pipette number
Enter the number of the pipette. Press the
Enter
button. The display ad-
vances to PIPETTE SETUP with NOMINAL highlighted.
Enter
PIPETTE SETUP
LIBRARY NAME
P2
TEST LIQUID(G/CC)
0.9971
BARO PRESSURE
1.0ATM
AUTO SAMPLE
OFF
AUTO PRINT
OFF
NUMBER OF SAMPLES
10
PIPETTE NAME
PIPETTE NAME
Enter pipette name
Enter the name of the pipette. Press
Enter
button. Display advances to
PIPETTE NUMBER. Press
Enter
button.
WHEAT
PIPETTE NUMBER
345
Select nominal
Press the
Enter
button. The display advances to NOMINAL VALUE UNITS with
ML highlighted.
Enter
PIPETTE SETUP
BARO PRESSURE
1.0ATM
AUTO SAMPLE
OFF
AUTO PRINT
OFF
NUMBER OF SAMPLES
10
PIPETTE NAME
WHEAT
PIPETTE NUMBER
345
NOMINAL
0.00ML
Select nominal value unit
Select either ML (milliliters) or UL (microliters). Press the
Enter
button.
Display advances to NOMINAL.
NOMINAL VALUE UNITS
ML
UL
RETURN TO SETUP
or
Balance modes