Place the first sample from the pipette into the vessel and press
Enter
. The
display indicates SAMPLE #1. Place the second sample from the pipette into
the vessel and press
Enter
.
When AUTO SAMPLE has been selected, it is not necessary to press the
Enter
button for each sample. If the balance fails to advance to the next
sample, press the
Enter
button.
Repeat the above steps until all samples have been run. When the last
sample is placed on the balance and the
Enter
button is pressed, the panel
display indicates the results of the test which includes the Date, Time, Status,
Nominal value, Inaccuracy and Imprecision. The graph on the display
indicates the mean value in the center with standard deviation and
2
times standard deviation above and below shown as lines across the
screen. Each sample is shown as a diamond shaped mark on the graph. A
ten sample set is shown.
If AUTO PRINT was selected during the setup, the results are automatically
printed. Printout contains numerical results and all statistics.
For manual printing of test results, press the
button on the balance.
P1
STATUS PASS
05/26/99 03:11:03P INACCURACY: 0.30%
NOMINAL: 3.45ML IMPRECISION:0.97%
3.525 2S
3.492 S
3.458 M
3.425 S
3.391 2S