3.3.1 Auto Calibration (AutoCal
TM
)
On Voyager balances equipped with the AutoCal
TM
feature, calibration can be accomplished using an internal calibra-
tion mass. Auto calibration can be performed at any time providing the balance has warmed up to operating tempera-
ture.
Go Back
Help
Enter
NOTE:
AutoCal
TM
calibration uses an internal mass in the
balance for calibration and is done automatically when
selected.
• Press
arrow button to select MAIN MENU.
• Press
Enter
button.
• Using the arrow buttons, select CALIBRATION.
• Press
Enter
button.
• Press
or
arrow button and select AUTOCAL.
• Clear the pan.
• Press
Enter
button, screen displays PLEASE WAIT,
the balance is automatically calibrated and then re-
turns to a weighing mode.
PROCEDURE
MAIN MENU
CHANGE UNITS
SET BALANCE
RETURN TO
WEIGH
CUSTOM MENU
CHANGE MODE
CALIBRATION
MAIN MENU
CHANGE UNITS
SET BALANCE
LIBRARY
CUSTOM MENU
CHANGE MODE
CALIBRATION
CALIBRATE
SPAN
LINEARITY
USER
CALIBRATION TEST
AUTOCAL
EXIT TO WEIGH
MAIN MENU
CHANGE UNITS
SET BALANCE
RETURN TO
WEIGH
CUSTOM MENU
CHANGE MODE
CALIBRATION
CALIBRATE
SPAN
LINPLEASEERARITY
USER
CALIBRATION TEST
AUTOCAL
EXIT TO WEIGH
INTERNAL CALIBRATION
PLEASE CLEAR THE PAN
AND PRESS < ENTER > . . .
MAIN MENU
CHANGE UNITS
SET BALANCE
LIBRARY
CUSTOM MENU
CHANGE MODE
CALIBRATION
CALIBRATE
SPAN
LINPLEASEERARITY
USER
CALIBRATION TEST
AUTOCAL
EXIT TO WEIGH
INTERNAL CALIBRATION
PLEASE WAIT . . .
0.00
WEIGHT
0%
100%
GRAMS
STABLE
MOVE HIGHLIGHT
MAIN MENU
CONTRAST
0.00
WEIGHT
0%
100%
GRAMS
STABLE
MOVE HIGHLIGHT
MAIN MENU
CONTRAST