71
7.13 Legal for trade (LFT)
LFT is a software controlled option which can be set to Legal For Trade ON or OFF. When set ON, certain items in the
CALIBRATION and SET BALANCE menus are automatically preset and locked to permit the balance to operate in a
legal for trade application and works in conjunction with a
Lock Out
switch (see section 7.14). Default setting is OFF.
The READOUT menu enables you to set the balance legal for trade (LFT) ON or OFF. See default table.
Enter
SET BALANCE
READOUT
From main menu
Enter SET BALANCE and select READOUT. The READOUT SETUP menu is
displayed.
READOUT SETUP
LEGAL FOR TRADE
Select LEGAL FOR TRADE, press
Enter
button .
Legal for trade default settings
The following table indicates the menus and options which are locked and
unlocked when LFT is set ON.
LFT DEFAULT TABLE
LFT and Lockswitch
Default Value
Set Balance Menu
Readout
Averaging Level
Unlocked
Stability Level
Locked to 1
Auto Zero
Limited to OFF & 0.5
Interface
Unlocked
Print Option
Auto Print
Unlocked
Stable Data
Locked ON
Numeric Data
Unlocked
Print Data
Unlocked
Print Time
Unlocked
Print Reference
Unlocked
Setup GLP
Unlocked
Set Time and Date
Unlocked
Autocal Enable
Locked
Print Current Settings
Unlocked
When LFT is turned ON, the last digit on the weight display has a
white colored block behind it. This signifies that the balance is in a
legal for trade operational mode and that the last digit should be
ignored. The CENTER OF ZERO is displayed only for LFT operation.
Balance settings
Summary of Contents for Voyager V10640
Page 61: ...55...
Page 62: ...56...
Page 65: ...1 Ohaus Corporation 29 Hanover Road Florham Park NJ 07932 0900 Operating instructions balances...
Page 154: ...Operating instructions balances O T O T O T O T OHAUS...
Page 243: ......
Page 245: ......