10
Go Back
Help
Enter
• Press
arrow button to select MAIN MENU.
• Press
Enter
button.
• Using the arrow buttons, select CALIBRATION.
• Press
Enter
button.
• Press
or
arrow button and select USER.
• Press
Enter
button, USER CALIBRATION is dis-
played.
• Using the arrow buttons, enter a value which is at
least 25% of the full span value, press
Enter
button.
The sample displays shown used a 2kg mass on an
4.1kg balance.
• Clear the pan and press
Enter
button.
• Place the selected mass value on the pan and press
Enter
button.
• Display indicates if calibration was successful and the
difference between the last calibration.
• Press any button, balance returns to weighing mode.
PROCEDURE
Go Back
Help
Enter
MAIN MENU
CHANGE UNITS
SET BALANCE
RETURN TO
WEIGH
CUSTOM MENU
CHANGE MODE
CALIBRATION
MAIN MENU
CHANGE UNITS
SET BALANCE
LIBRARY
CUSTOM MENU
CHANGE MODE
CALIBRATION
CALIBRATE
SPAN
LINEARITY
USER
CALIBRATION TEST
AUTOCAL
EXIT TO WEIGH
MAIN MENU
CHANGE UNITS
SET BALANCE
RETURN TO
WEIGH
CUSTOM MENU
CHANGE MODE
CALIBRATION
CALIBRATE
SPAN
LINEARITY
USER
CALIBRATION TEST
AUTOCAL
EXIT TO WEIGH
ENTER CAL VALUE
UP/DOWN TO SCROLL CHARACTERS
LEFT/RIGHT TO MOVE CURSOR
ENTER TO ACCEPT
BACK TO QUIT
2000
MAIN MENU
CHANGE UNITS
SET BALANCE
RETURN TO
WEIGH
CUSTOM MENU
CHANGE MODE
CALIBRATION
CALIBRATE
SPAN
LINEARITY
USER
CALIBRATION TEST
AUTOCAL
EXIT TO WEIGH
ENTER CAL VALUE
UP/DOWN TO SCROLL CHARACTERS
LEFT/RIGHT TO MOVE CURSOR
ENTER TO ACCEPT
BACK TO QUIT
0
USER CALIBRATION
PLEASE CLEAR THE PAN
AND PRESS ENTER KEY . . .
MAIN MENU
CHANGE UNITS
SET BALANCE
LIBRARY
CUSTOM MENU
CHANGE MODE
CALIBRATION
CALIBRATE
SPAN
LINEARITY
USER
CALIBRATION TEST
AUTOCAL
EXIT TO WEIGH
ENTER CAL VALUE
UP/DOWN TO SCROLL CHARACTERS
LEFT/RIGHT TO MOVE CURSOR
ENTER TO ACCEPT
BACK TO QUIT
0
USER CALIBRATION
PLEASE PUT: 2000 G ON PAN.
AND PRESS < ENTER > . . .
MAIN MENU
CHANGE UNITS
SET BALANCE
RETURN TO
WEIGH
CUSTOM MENU
CHANGE MODE
CALIBRATION
CALIBRATE
SPAN
LINEARITY
USER
CALIBRATION TEST
AUTOCAL
EXIT TO WEIGH
ENTER CAL VALUE
UP/DOWN TO SCROLL CHARACTERS
LEFT/RIGHT TO MOVE CURSOR
ENTER TO ACCEPT
BACK TO QUIT
0
USER CALIBRATION
PLEASE WAIT . . .
MAIN MENU
CHANGE UNITS
SET BALANCE
LIBRARY
CUSTOM MENU
CHANGE MODE
CALIBRATION
CALIBRATE
SPAN
LINEARITY
USER
CALIBRATION TEST
AUTOCAL
EXIT TO WEIGH
ENTER CAL VALUE
UP/DOWN TO SCROLL CHARACTERS
LEFT/RIGHT TO MOVE CURSOR
ENTER TO ACCEPT
BACK TO QUIT
0
USER CALIBRATION
USER CAL SUCCESSFULLY DONE!
THE LAST DIFFERENCE BETWEEN THIS CAL
AND LAST CAL IS:
0.01 GRAMS
PRESS ANY KEY TO EXIT . . .
3.3.4 User Calibration
User calibration is used when it is desired to calibrate the balance using a mass of known value. To use this calibra-
tion feature, proceed as follows:
0.00
WEIGHT
0%
100%
GRAMS
STABLE
MOVE HIGHLIGHT
MAIN MENU
CONTRAST
0.00
WEIGHT
0%
100%
GRAMS
STABLE
MOVE HIGHLIGHT
MAIN MENU
CONTRAST
Summary of Contents for Voyager V10640
Page 61: ...55...
Page 62: ...56...
Page 65: ...1 Ohaus Corporation 29 Hanover Road Florham Park NJ 07932 0900 Operating instructions balances...
Page 154: ...Operating instructions balances O T O T O T O T OHAUS...
Page 243: ......
Page 245: ......