12
Installation
2.2 Selecting the location
The balance should always be used in an environment which is free from excessive air currents, corrosives,
vibration, and temperature or humidity extremes. These factors will affect displayed weight readings.
!"
#
>O/T<
>O/T<
>O/T<
>O/T<
!"
#
!"
#
>O/T<
>O/T<
DO NOT
install the balance:
Next to open windows or doors causing drafts or
rapid temperature changes.
Near air conditioning or heat vents.
Near vibrating, rotating or reciprocating equipment.
Near magnetic fields or equipment that generates
magnetic fields.
On an unlevel work surface.
In confined areas, allow sufficient space around the
instrument for ease of operation and keep away from
radiating heat sources.
2.3 Leveling the balance
Exact horizontal positioning and stable installation are prerequisites for repeatable results. To compensate for small
irregularities or inclinations at the location, the instrument can be leveled.
Adjust the leveling feet at the rear of the balance until the air bubble in the
indicator is centered.
NOTE
: The instrument should be leveled each time its location is changed.
>O/T<
>O/T<
OHA
US
®
! "#
$
Leveling
Indicator
Leveling
Foot
Balances are shipped with the pan not installed. On balances equipped with
a draft shield, slide open the side door and insert the pan into the center
hole which is the measuring transducer.
2.4 Installing pan
>O/T<
>O/T<
>O/T<
>O/T<
OHA
US
®
0.000
BASIC
WEIG
HING
0%
100%
GRAM
S
STABL
E
Enter
Go B
ack
Help
PRESS <ENTER> FO
R M
ENU
Pan
Summary of Contents for Voyager V10640
Page 61: ...55...
Page 62: ...56...
Page 65: ...1 Ohaus Corporation 29 Hanover Road Florham Park NJ 07932 0900 Operating instructions balances...
Page 154: ...Operating instructions balances O T O T O T O T OHAUS...
Page 243: ......
Page 245: ......