20
Calibration
4.1 Span calibration
Span calibration normally requires that calibration be made using a mass equal to the full capacity of the balance,
however, the Voyager balance can be calibrated using other lesser values as specified on the display.
CALIBRATE
SPAN
LINEARITY
USER
CALIBRATION TEST
AUTOCAL
AUTOCAL DELTA CORRECTION
EXIT TO WEIGH
From main menu
Scroll to CALIBRATION and press the
Enter
button. Display advances to
CALIBRATE with SPAN highlighted.
Enter
SPAN CALIBRATION
PLEASE CLEAR THE PAN. . .
AND PRESS <ENTER>.
Clear the pan.
Enter
SPAN CALIBRATION
PLEASE PUT 400.000 GRAMS
ON PAN AND PRESS <ENTER...
OTHER WEIGHTS MAY BE USED:
300 200 100
Enter
Place indicated mass value on pan.
SPAN CALIBRATION
PLEASE WAIT. . .
THE DIFFERENCE BETWEEN THIS CAL
AND THE LAST CAL IS: 0.000 GRAMS
PRESS <ENTER> TO CONTINUE
Enter
Display indicates the difference between the present calibration and the
last calibration.
400.000
GRAMS STABLE
BASIC WEIGHING
Enter
PRESS <ENTER> FOR MENU
0% 100%
Display indicates weight of the calibration masses. Remove masses from
pan. Balance returns to a weighing mode.
Summary of Contents for Voyager V10640
Page 61: ...55...
Page 62: ...56...
Page 65: ...1 Ohaus Corporation 29 Hanover Road Florham Park NJ 07932 0900 Operating instructions balances...
Page 154: ...Operating instructions balances O T O T O T O T OHAUS...
Page 243: ......
Page 245: ......