Calibration
4.2Linearity calibration
Linearity calibration utilizes three calibration points, one at zero, center span and full span. This method minimizes
deviation between actual and displayed weights within the balance's weighing range. Three weight values are used;
zero, a weight value at midpoint of the balance's weighing range and a weight value at or near the specified capacity.
CALIBRATE
SPAN
LINEARITY
USER
CALIBRATION TEST
AUTOCAL
AUTOCAL DELTA CORRECTION
EXIT TO WEIGH
From main menu
Scroll to CALIBRATION and press the
Enter
button. Display advances to
CALIBRATE with SPAN highlighted. Scroll to LINEARITY.
Enter
LINEARITY CALIBRATION
PLEASE CLEAR THE PAN. . .
AND PRESS <ENTER>.
Clear the pan.
Enter
LINEARITY CALIBRATION
PLEASE PUT 200.000 GRAMS
ON PAN AND PRESS <ENTER...
Enter
Place indicated mass value on pan.
LINEARITY CALIBRATION
PLEASE WAIT. . .
THE DIFFERENCE BETWEEN THIS CAL
AND THE LAST CAL IS: 0.000 GRAMS
PRESS <ENTER> TO CONTINUE
Enter
Display indicates the difference between the present calibration and the
last calibration.
400.000
GRAMS STABLE
BASIC WEIGHING
Enter
PRESS <ENTER> FOR MENU
0% 100%
Display indicates weight of the calibration masses. Remove masses from
pan. Balance returns to a weighing mode.
LINEARITY CALIBRATION
PLEASE PUT 400.000 GRAMS
ON PAN AND PRESS <ENTER...
Enter
Place indicated mass value on pan.