64
Balance settings
7.4 Setting the interface
This menu option enables you to set the balance communication parameters for the RS232 interface which
include: baud rate, data bits, parity and stop bit.
Setting baud rate
Baud rates of 300, 1200, 2400, 4800 and 9600 are available for commu-
nications. Select the appropriate rate. Default setting is 2400 baud. Press
Enter
button.
INTERFACE
BAUD RATE
or
300
1200
2400
4800
9600
RETURN
Setting data bits
Data bits of 7 or 8 are selectable. Select appropriate bit rate. Default setting
is 7. Press
Enter
button.
INTERFACE
DATA BITS
or
7
8
RETURN
Setting parity
Select appropriate parity setting of either NONE, EVEN, ODD, 0, or 1. Default
setting is NONE. Press
Enter
button.
INTERFACE
PARITY
or
NONE
EVEN
ODD
0
1
RETURN
Setting stop bits
Select appropriate stop bits setting of either 1 or 2. Default setting is 2.
Press
Enter
button.
INTERFACE
STOP BITS
or
1
2
RETURN
SET BALANCE
Enter
From main menu
Select SET BALANCE and press the
Enter
button. Display advances to SET
BALANCE with READOUT highlighted. Scroll to INTERFACE and press the
Enter
button, display advances to INTERFACE.
INTERACE