22
Calibration
4.3 User calibration
User calibration is used when it is desired to calibrate the balance using a mass of known value. To use this
calibration feature, proceed as follows:
CALIBRATE
SPAN
LINEARITY
USER
CALIBRATION TEST
AUTOCAL
AUTOCAL DELTA CORRECTION
EXIT TO WEIGH
From main menu
Scroll to CALIBRATION and press the
Enter
button. Display advances to
CALIBRATE with SPAN highlighted. Scroll to USER.
Enter
USER CALIBRATION
USER CALIBRATION RANGE:
100-400 GRAMS
PRESS <ENTER. TO CONTINUE.
User calibration range is specified.
Enter
USER CALIBRATION
PLEASE PUT 200.000 GRAMS
ON PAN AND PRESS <ENTER...
Enter
Place selected mass value on pan.
USER CALIBRATION
PLEASE WAIT. . .
THE DIFFERENCE BETWEEN THIS CAL
AND THE LAST CAL IS: 0.000 GRAMS
PRESS <ENTER> TO CONTINUE
Enter
Display indicates the difference between the present calibration and the
last calibration.
200.000
BASIC WEIGHING
Enter
0% 100%
Display indicates weight of the calibration masses. Remove masses from
pan. Balance returns to a weighing mode.
USER CALIBRATION
ENTER CAL VALUE IN GRAMS
200.000
200.000
Enter the calibration value.
GRAMS STABLE
PRESS <ENTER> FOR MENU
USER CALIBRATION
PLEASE CLEAR THE PAN. . .
AND PRESS <ENTER>.
Clear the pan.
Enter