47
Balance modes
AUX LIQ & MODE
H20
<-SELECTED
OTHER
POROUS MATERIAL
OFF
DRY WEIGHT
O.000
DENSITY OF OIL
0.0000
RETURN TO SETUP
Enter
Selecting water as auxiliary liquid
Select H2O, and perform the next step. If you are using a different liquid, skip
the next step and continue.
AUX LIQ & MODE
ENTER TEMPERATURE CELSIUS
Specify temperature
Enter the temperature of the water and press
Enter
button. Display
advances to AUX LIQ & MODE with POROUS MATERIAL highlighted. Skip the
next two steps.
AUX LIQ & MODE
H20
<-SELECTED
OTHER
POROUS MATERIAL
OFF
DRY WEIGHT
O.000
DENSITY OF OIL
0.0000
RETURN TO SETUP
Enter
Selecting an auxiliary liquid
Select OTHER, and perform the next step.
AUX LIQ & MODE
ENTER LIQUID DENSITY IN CC
Specify auxiliary liquid density
Enter the density of the auxiliary liquid and press
Enter
button. Display
advances to AUX LIQ & MODE with POROUS MATERIAL highlighted.
AUX LIQ & MODE
H20
<-SELECTED
OTHER
POROUS MATERIAL
OFF
DRY WEIGHT
O.000
DENSITY OF OIL
0.0000
RETURN TO SETUP
Enter
Return to setup
Select RETURN TO SETUP. Disregard POROUS MATERIAL in menu as this is
explained in Appendix A-2. Display returns to DENSITY SETUP with AUTO
SAMPLE highlighted.
DENSITY SETUP
Enter
LIBRARY NAME
D2
SOLID DENSITY
<-SELECTED
LIQUID DENSITY
AUTO SAMPLE
OFF
AUTO PRINT
OFF
RUN
Auto sample
Select AUTO SAMPLE and turn it ON or OFF. A setting of ON allows samples
to be sequentially sampled without pressing the
Enter
button for each
sample.
DENSITY SETUP
Enter
LIBRARY NAME
D2
SOLID DENSITY
<-SELECTED
LIQUID DENSITY
AUTO SAMPLE
OFF
AUTO PRINT
OFF
RUN
Auto print
At this point you can select AUTO PRINT. To turn on, press the
Enter
button
and select ON. If you do not want automatic printing, select OFF.
DENSITY SETUP
Enter
LIBRARY NAME
D2
SOLID DENSITY
<-SELECTED
LIQUID DENSITY
AUTO SAMPLE
OFF
AUTO PRINT
OFF
RUN
Testing the sample
Select RUN and press the
Enter
button. The display advances to DENSITY
and requests WEIGH IN AIR shown at the bottom of the screen.
0.000
START/SAMPLE SETUP MAIN MENU
DENSITY - - D2
WEIGH IN AIR: 0.000
TEMP:25.0
WEIGH IN LIQ:
AUX D= 9.971
VOLUME:
WEIGH IN AIR