60
Balance modes
Enter nominal value
Enter the nominal value. Press the
Enter
button. The display advances to
PIPETTE SETUP with INACCURACY highlighted.
Select inaccuracy
Select INACCURACY. Press
Enter
button. The display advances to INACCU-
RACY.
NOMINAL
0.0000
0.0000
Enter
PIPETTE SETUP
AUTO SAMPLE
OFF
AUTO PRINT
OFF
NUMBER OF SAMPLES
10
PIPETTE NAME
WHEAT
PIPETTE NUMBER
345
NOMINAL
3.45ML
INACCURACY
2.00%
Select imprecision
Select IMPRECISION. Press the
Enter
button. Display advances to IMPRECI-
SION.
Enter imprecision value
Enter the imprecision value. This indicates the variation of all pipette
samples from the mean value (Coefficient of Variation). Press the
Enter
button. The default setting is 2 percent. The display advances to RUN.
PIPETTE SETUP
AUTO PRINT
OFF
NUMBER OF SAMPLES
10
PIPETTE NAME
WHEAT
PIPETTE NUMBER
345
NOMINAL
3.45ML
INACCURACY
2.00%
IMPRECISION
2.00%
INACCURACY
Enter inaccuracy value
Enter the percentage of acceptable inaccuracy. This indicates the deviation of
all pipette samples (mean error). Press
Enter
button. The default setting is 2
percent. Display advances to IMPRECISION.
2.000
2.000
Enter
IMPRECISION
2.000
2.000
When placing samples in the vessel during pipette calibration, make sure to
continue the calibration to the end without significant pauses in between
each sample. If a significant amount of time has elapsed (30 seconds or
more) between samples, re-tare the balance before adding the next sample.
This is to eliminate errors introduced by evaporation.
Enter
PIPETTE SETUP
PIPETTE NUMBER
345
NOMINAL
3.45ML
INACCURACY
2.00%
IMPRECISION
2.00%
VIEW RESULTS
CLEAR ALL DATA
RUN
0.000
AUTO SAMPLE ON GRAMS STABLE
PIPETTE - - P2
Taking samples
With RUN highlighted, press the
Enter
button. The display advances to a
screen requesting to PUT SAMPLE #1 ON PAN.
0%
100%
PUT SAMPLE #1 ON PAN
Press
Enter
, the display indicates PIPETTE -- NAME of test at the top of the
screen.
START/SAMPLE is at the bottom of the screen. Pressing
Enter
starts the
testing. If you are using water, use distilled water only.
Place empty vessel on the balance, fill the vessel with a small amount of
liquid, and then tare the balance.
START/SAMPLE SETUP MAIN MENU