78
RS232 Interface/printing data
Pipette test printout
When performing a Pipette Test with GLP turned on, a printout is available.
The following sample printout is shown.
Pipette Results
Library Name: WHEATON
Pipette Name: WHEATON
Pipette Number: 832
Water Temp(C): 25.0
Test Liquid Density(g/cc): 0.9971
Barometric Pressure: 1.00PSIA
Nominal Value: 3.448ml
INACCURACY
E%: 0.31%
E% Limit: 2.00%
Mean Value: 3.459ml
IMPRECISION
CV: 0.88%
CV Limit: 2.00%
Standard Deviation: 0.031ml
Status: PASS
07/01/99
-----Sample Data ml----
3.488
3.485
3.487
3.435
3.458
3.430
3.470
3.510
3.446
3.456
Operator:----------------------