33
3.16 Printing Data (Cont.)
Span Calibration Printout
When performing a Span calibration, a printout is auto-
matically made after the calibration mass is placed on
the pan and the
button is pressed.
Calibration Test Printout
When performing a Calibration Test with GLP turned on,
a printout is available. When the display indicates the
mass value to be placed on the pan, the balance the
automatically displays the calibration weight required
and the
button is pressed.
Linearity Calibration Printout
When performing a Linearity calibration with GLP turned
on, a printout is automatically made after the calibration
mass is placed on the pan and the
button is
pressed.
- - - - - CAL TEST - - - - - -
7/01/96
1:00:00 PM
Bal Id 1234
Cal:
4000.00g
Act:
4000.04g
Dif:
0.04g
Wt. Ref......................................
USER NO 2056853
PROJ NO 100012
Name........................................
- - - - - END - - - - -
- - - - - SPAN CAL - - - - - -
7/01/96
1:00:00 PM
Bal Id 1234
Cal:
4000.00g
Old:
4000.00g
Dif:
0.00g
Wt. Ref......................................
USER NO 2056853
PROJ NO 100012
Name........................................
- - - - - END - - - - -
- - - - - LIN CAL - - - - - -
7/01/96
1:00:00 PM
Bal Id 1234
Cal:
4000.00g
Old:
3999.94g
Dif:
0.06g
Wt. Ref......................................
USER NO 2056853
PROJ NO 100012
Name........................................
- - - - - END - - - - -