13
Wind Shield
On 610g to 4100g balances with 0.01g resolution, a wind shield is
required to reduce the possibility of air currents from disturbing the pan.
Make sure the wind shield is firmly snapped into place.
2.5 Installing wind shield
2.6 Power supply
Check to ensure the voltage printed on the AC adapter matches your local
line voltage. If this is not the case, on no account connect the AC adapter to
the power supply, but contact your responsible OHAUS dealer.
Rear of Balance
AC Adapter Connection
Connect the AC Adapter supplied to the three pin connector located at the
rear of the balance.
The balance now performs a self-test, loads parameters, displays the
software version, capacity, indicates if autocal is installed and the normal
weight display then appears.
Ensure that the AC adapter can never come into contact with liquids!
Allow your balance to warm up for at least 30 minutes
to enable it to
adapt itself to the ambient conditions. If the balance has been stored in a
very cold environment before installation, it may require several hours for the
balance to stabilize.