86
A-2
Porous material density determinations
The density of a porous (oil impregnated part) can be made with the balance. Weigh the part (dry) prior to oil
impregnation and
record its weight
. You must also know the density value of the oil to be used in immersing the
part before starting. In this procedure, you will follow the method for solid density measurements using water.
From main menu
Enter CHANGE MODE. Select DENSITY, display advances to LIBRARY NAME.
MAIN MENU
CHANGE MODE
Enter
DENSITY SETUP
LIBRARY NAME
Library name
Enter a library name, up to 8 characters and press
Enter
button. Display
advances to DENSITY SETUP with SOLID DENSITY highlighted. If a previous
name was entered and selected, the balance uses the parameters which
were previously set.
DENSITY SETUP
Enter
LIBRARY NAME
D2
SOLID DENSITY
<-SELECTED
LIQUID DENSITY
AUTO SAMPLE
OFF
AUTO PRINT
OFF
RUN
Select SOLID DENSITY, display advances to AUX LIQ & MODE with H2O
highlighted.
AUX LIQ & MODE
H20
<-SELECTED
OTHER
POROUS MATERIAL
OFF
DRY WEIGHT
O.000
DENSITY OF OIL
0.0000
RETURN TO SETUP
Enter
Selecting water as auxiliary liquid
Select H2O, display advances to AUX LIQ & MODE with ENTER TEMPERA-
TURE CELSIUS highlighted.
AUX LIQ & MODE
ENTER TEMPERATURE CELSIUS
Specify temperature
Enter the temperature of the water and press
Enter
button. Display
advances to AUX LIQ & MODE with POROUS MATERIAL highlighted.
AUX LIQ & MODE
Specify porous material
Select Porous Material and press
Enter
button. Display advances to POROUS
MATERIAL ON/OFF. Select ON and press
Enter
button.
H20
<-SELECTED
OTHER
POROUS MATERIAL OFF
DRY WEIGHT
O.000
DENSITY OF OIL
0.0000
RETURN TO SETUP
Enter
AUX LIQ & MODE
H20
<-SELECTED
OTHER
POROUS MATERIAL ON
DRY WEIGHT
200.040
DENSITY OF OIL
0.0000
RETURN TO SETUP
Enter dry weight
Enter the dry weight of the part as previously measured. Press
Enter
button.
Display advances to DENSITY OF OIL highlighted.
Appendix A - special applications
Summary of Contents for Voyager V10640
Page 61: ...55...
Page 62: ...56...
Page 65: ...1 Ohaus Corporation 29 Hanover Road Florham Park NJ 07932 0900 Operating instructions balances...
Page 154: ...Operating instructions balances O T O T O T O T OHAUS...
Page 243: ......
Page 245: ......