25
Calibration
4.6 AutoCal
TM
delta correction
Balances with AutoCal
TM
contain software which allows the internal calibration mass to be ad 100
divisions at full scale capacity. This permits calibrating the balance using an external ASTM Class 1 mass which is
traceable to a certified standard.
CALIBRATE
SPAN
LINEARITY
USER
CALIBRATION TEST
AUTOCAL
AUTOCAL DELTA CORRECTION
EXIT TO WEIGH
From main menu
Scroll to CALIBRATION and press the
Enter
button. Display advances to
CALIBRATE with SPAN highlighted. Scroll to AUTOCAL DELTA CORRECTION.
Enter
INTERNAL CALIBRATION
PLEASE CLEAR THE PAN. . .
AND PRESS <ENTER>.
Clear the pan. Internal calibration starts.
INTERNAL CALIBRATION
PLEASE WAIT. . .
THE DIFFERENCE BETWEEN THIS CAL
AND THE LAST CAL IS: 0.000 GRAMS
PRESS <ENTER> TO CONTINUE
Display indicates if calibration was successful and the difference between
the last calibration. It is not necessary to press the
Enter
button, the balance
returns to a weighing mode automatically. If a difference still exists, repeat
the procedure. The difference should be 0.000 grams.
0.000
BASIC WEIGHING
0% 100%
Balance returns to a weighing mode.
GRAMS STABLE
PRESS <ENTER> FOR MENU
CALIBRATE
AUTOCAL DELTA CORRECTION
Enter the number of digits shown between the span calibration value of
the balance and the actual reading. For example, on a 400g balance, the
reading is 400.002g. The difference is 0.002g. This indicates a correction of
2 divisions is required.
0.000
BASIC WEIGHING
0% 100%
PRESS <ENTER> FOR MENU
Tare the balance, then place an ASTM Class 1 mass equal to the span
calibration value of the balance. Note the reading. If the reading is higher or
lower than the value of the mass, proceed as follows:
- > O / T < -
2
2
Enter