56
Balance modes
Entering current library name
The current name you assigned should be displayed. Press the
Enter
button.
Display advances to SQC STARTUP with RESUME highlighted.
6.10.3
Redoing or correcting a batch
If during the course of taking samples, an error has occurred and you want to rerun the set of samples you are
currently working with, this procedure will allow you to start a new set without recording data from the set containing
errors.
SQC SETUP
Go Back
LIBRARY NAME
TOLERANCE
TARE
MEAN TARE VALUE
AUTO SAMPLE
AUTO PRINT
N NUMBER OF SAMPLES
BATCH NAME
Exiting a sample set containing an error
When the current sample you are working with is incorrect, you can exit by
pressing the
Go Back
button. The display returns back to SQC SETUP with
LIBRARY NAME highlighted. Press the
Enter
button.
STATISTICS SETUP
LIBRARY NAME
SQC STARTUP
SQC SETUP
RESUME
NEW SETUP
PREVIOUS SETUP/REDO BATCH
PREVIOUS SETUP/CLEAR ALL DATA!
Enter
Enter
Select previous setup
Select PREVIOUS SETUP/REDO BATCH and press
Enter
button. The display
returns to SQC SETUP with VIEW BATCH HISTORY highlighted.
Select run
Scroll to RUN. With RUN selected, press the
Enter
button. The balance
returns to the sample set interrupted and you may continue to take samples.
MEAN TARE VALUE OFF
AUTO SAMPLE
OFF
AUTO PRINT
OFF
NUMBER OF SAMPLES
3
BATCH NAME
N01
VIEW MEAN VALUE TRACE
VIEW BATCH HISTORY
RUN
Clearing previous data set
When you are in the SQC STARTUP menu, you can elect to remove all data if
desired by selecting PREVIOUS SETUP/CLEAR ALL DATA and pressing
Enter
button.
Summary of Contents for Voyager V10640
Page 61: ...55...
Page 62: ...56...
Page 65: ...1 Ohaus Corporation 29 Hanover Road Florham Park NJ 07932 0900 Operating instructions balances...
Page 154: ...Operating instructions balances O T O T O T O T OHAUS...
Page 243: ......
Page 245: ......