Balance settings
Select any one of the available options and press
Enter
button.
Enter
From main menu
Enter the SET BALANCE menu and scroll to LOCK OUT.
SET BALANCE
LOCK OUT
7.10 Lock out
This software option works in conjunction with a hardware Lockswitch and LFT software. It permits various measure-
ment units, calibration methods and balance functions to be selected and either turned ON or OFF.
Before using this sofware function, check the status of the hardware lock switch and Legal for Trade software
(see section 7.14). If the hardware switch is ON, and the Legal for Trade software is set ON (locked), you cannot
access the LFT LOCK software.
LFT LOCK
or
TURN UNITS <ON> OR <OFF>
TURN CALIBRATION <ON> OR <OFF>
TURN FUNCTIONS <ON> OR <OFF>
Depending on what you have selected, an entire list of all Units, Calibration
methods and balance Functions are displayed. Every item in each list can be
individually turned ON or OFF. These items when turned OFF will not be
available for balance operation and are locked out.
All selected items which have been turned OFF will still function. Menu lock
out is accomplished with the setting of a physical switch located under the
pan. Paragraph 7.14 describes how to lock out the selected menus. The
balance can then be sealed if required.