63
Balance settings
7.3 Setting the readout
This menu option enables you to set the balance averaging level, stability level, auto zero, and legal for trade
(LFT).
Setting averaging level
Press
Enter
button, AVERAGING LEVEL FILTER is displayed. Averaging level
filter compensates for vibration or excessive air currents. A zero setting
(GOOD) corresponds to minimum filtering with fastest stabilization time. A
setting of (BEST)10 is maximum filtering with the slowest stabilization time.
Adjust to desired setting and press
Enter
button.
Setting stability level
Scroll to STABILITY LEVEL and press
Enter
button, STABILITY LEVEL
FILTER is displayed. Stability level filter determines the variation range in
divisions for a given reading depending upon the filter setting. A setting of
zero (GOOD) permits the balance to respond to variations of .5 divisions. A
setting of (BEST) 10 sets the threshold to 3 divisions. Adjust to desired
setting and press
Enter
button.
READOUT SETUP
AVERAGING LEVEL FILTER
GOOD
BEST
5
or
READOUT SETUP
STABILITY LEVEL FILTER
GOOD
BEST
5
or
Setting auto-zero
Scroll to AUTO ZERO and press
Enter
button, AZT is displayed. Auto zero
minimizes the effects of temperature changes and shift on the zero reading.
There are 4 settings, OFF, 0.5, 1, and 3. The numbers represent display
divisions. The balance maintains the zero display until the set threshold is
exceeded. Select the desired setting and press
Enter
button.
READOUT SETUP
AZT
or
OFF
0.5
1
3
Setting legal for trade
Scroll to LEGAL FOR TRADE and press
Enter
button. Legal for trade can be
set ON or OFF. When set ON, only the functions allowed by the national
weights and measures legislation are available. Press Enter button after
making your selection.
READOUT SETUP
AVERAGING LEVEL 5
STABILITY LEVEL
5
AUTO ZERO 0.5
LEGAL FOR TRADE OFF
EXIT TO WEIGH
or
SET BALANCE
Enter
From main menu
Select SET BALANCE and press the
Enter
button. Display advances to SET
BALANCE with READOUT highlighted. Press the
Enter
button, display
advances to READOUT SETUP.
READOUT
Summary of Contents for Voyager V10640
Page 61: ...55...
Page 62: ...56...
Page 65: ...1 Ohaus Corporation 29 Hanover Road Florham Park NJ 07932 0900 Operating instructions balances...
Page 154: ...Operating instructions balances O T O T O T O T OHAUS...
Page 243: ......
Page 245: ......