52
5.3 Error Codes List
Error Codes List
The following list describes the various error codes and which can appear on the display and the suggested remedy.
Data Errors
1.0
Transient error (hardware error, probably static discharge). If error persists, the balance must be
serviced.
1.1
Balance temperature transducer hardware error.
1.2
No data from main board.
Tare Errors
2.0
Balance is unable to stabilize within time limit after taring. Environment is too hostile or balance needs
recalibration.
Calibration Errors
3.0
Incorrect or no calibration mass used for calibration. Recalibrate with correct masses.
3.1
Balance was unable to acquire stable data during calibration. The balance will try again.
RS232 Errors
4.4
RS232 buffer is full.
User Errors
7.0
User entry out of bounds.
7.2
Number outside of display capacity.
Over-Under Load Errors
8.0
Hardware error causing an internal weight signal which is too low. Check if pan is off. If not, the balance
must be serviced.
8.1
Hardware error caused by an internal weight signal which is too high. Check load on the pan which may
be excessive. If error persists, the balance must be serviced.
8.2
Power-on load out of specification (LFT only)
8.3
Rated capacity exceeded. Remove excessive weight from pan.
8.4
Underload condition on balance. Check that the proper pan is installed.
8.5
AutoCal
TM
weight internal sensor indicated its weight on the pan.
CheckSum Errors
9.1
Bad factory checksum. If error persists, have the balance serviced.
9.2
Bad factory checksum. If error persists, have the balance serviced.
9.3
Bad factory checksum. If error persists, have the balance serviced.
9.4
AutoCal
TM
data failed checksum. This failure will disable access to the autocal feature (if installed).
9.5
Factory calibration data failed checksum.
9.6
Bad program checksum.
9.7
Bad CMOS checksum.
9.8
User calibration data failed checksum.
9.9
Temperature compensation data failed checksum.
5.4 Information Messages
Informational messages such as calibrate now and error messages appear on the display either when an action is
required on the users part or a malfunction has occurred in the balance due to hardware, software errors or misappli-