54
Capacity (g)
62
110
210
100/210 *
Readability (mg)
0.1
0.1/1
Weighing modes
g, mg, oz , oz t, ct, dwt, taels (3), mommes, gn, ti, N, custom unit
Functions
Parts counting, differential weighing, quick check, statistics, formulation, filling, animal weighing, percent
formulation, check weighing, density determination, SQC & pipette calibration.
Performance enhancing features
Program-Link
TM,
GLP protocol, contrast and brightness control, custom menu, pop-up windows, application
library, go-back key, help text
Repeatability (Std. dev.) (mg)
0.1
0.1/0.5
Linearity (mg) (+)
0.2
0.2/0.5
Tare range
Full capacity by subtraction
Safe overload capacity
150% of capacity
Stabilization time
<4 seconds
Sensitivity drift PPM/°C
3
(10°C - 30°C)
Operating temperature range:
w/internal calibration
10° to 40°C/ 50° to 104°F
w/o internal calibration
10° to 30°C/50° to 86°F
Calibration
Auto calibration
Power requirements
External Adapter, 100 -120 V ac, 220 - 240 V ac, 50/60 Hz
Plug configuration for US, Euro, UK, Japan & Australia
Draft shield (in/cm)
10.2/25.9
(free height above platform)
Display (in/cm)
0.6/1.5
Pan size (in/cm)
3.5/9. diameter
Dimensions
9 x 15.25 x14/22.8 x 38.7 x 35.5
(WxHxD) (in/cm)
5.8 Specifications
Analyticals
Capacity (g)
210
410
100/410* 610 2100 4100 1000/4100* 4100 6100 8100
Readability (g)
0.001
0.001/0.01
0.01
0.01/0.1
0.1
Weighing modes
g, mg, kg, lbs, oz , oz t, ct, dwt, taels (3), mommes, gn, ti, N, custom unit
Functions
Parts counting, differential weighing, quick check, statistics, formulation, filling, animal weighing, percent
formulation, check weighing, density determination, SQC & pipette calibration.
Performance enhancing features
Program-Link
TM,
GLP protocol, contrast and brightness control, custom menu, pop-up windows, application
library, go-back key, help text
Repeatability (Std. dev.) (g)
0.0005
0.0005/0.005
0.005 0.01/0.05 0.05
Linearity (g) (+)
0.002
0.002/0.005
0.02 0.02/0.05 0.1
Tare range
Full capacity by subtraction
Stabilization time
<3 seconds
Sensitivity drift PPM/°C
4
3
4
3
4 3 4 3
(10°C - 30°C)
Operating temperature range:
w/internal calibration
10° to 40°C/50° to 104°F
w/o internal calibration
10° to 30°C/50° to 86°F
Calibration
Auto-calibration
Power requirements
External Adapter, 100 -120 V ac, 220 - 240 V ac, 50/60 Hz
Plug configuration for US, Euro, UK, Japan & Australia
Draft shield (in/cm)
10.2/25.9
(free height above platform)
Display (in/cm)
0.6/1.5
Pan size (in/cm)
4.7/12 Dia.
6.8 x 6.8/17.2 x 17.2 w/windshield
8 x 8/ 20.3 x 20.3
Dimensions
9 x 15.25 x14/22.8 x 38.7 x 35.5
8.25 x 4 x 14/23.4 x 10.1 x 35.5
(WxHxD) (in/cm)
Precision Top Loaders
NOTE:
Not all weighing modes apply depending upon capacity and resolution of the balance.
* Moveable FineRange
TM
Summary of Contents for Voyager V10640
Page 61: ...55...
Page 62: ...56...
Page 65: ...1 Ohaus Corporation 29 Hanover Road Florham Park NJ 07932 0900 Operating instructions balances...
Page 154: ...Operating instructions balances O T O T O T O T OHAUS...
Page 243: ......