8 8700-0009-000
(Rev. 14) 11/19
Setting the suction level
1. Move the mode selector switch to | (CONT).
Operation
DISS fitting
1. Insert the safety trap into the regulator fitting. Situate the nipple in the
desired position.
2. Turn the DISS wing nut clockwise to engage threads and tighten (there is
no O-ring, so the vacuum seal depends on a tight connection).
Regulator
Wing nut
Safety Trap
Mode Selection
|O|O
(INT)
- suction is intermittent (cycled “ON” and “OFF”) and the
suction level can be adjusted with the suction control knob when cycled
“ON”.
O
(OFF)
- No suction is supplied to the patient.
I
(CONT)
- Suction is continuous and can be adjusted with the suction
control knob.
760
MA
X
100
MA
X
2. Push and rotate the suction control knob until the vacuum gauge indicates the required setting.
CAUTION:
The suction control knob must be completely pushed in to adjust the vacuum level. Failure to do
so may damage the vacuum regulator.