Warnings

Before using the Ohmeda Medical Panda

TM

 Baby Warmer System, read this

entire manual. Attempting to use this device without a thorough understanding
of its operation may result in patient or user injury. This device should only be
operated by personnel trained in its operation and under the direction of qualified
medical personnel familiar with the risks and benefits of this type of device.
Additional precautions specific to certain procedures are found in the
text of this manual.

Do not use the Warmer in the presence of flammable anesthetics; a possible
explosion hazard exists under these conditions.

Do not perform the Checkout Procedures (Mechanical and Control Unit) while a
patient occupies the Warmer.

Complete the “Checkout Procedures” section of this manual before putting the
unit into operation. If the warmer fails any portion of the checkout procedures it
must be removed from use and repaired.

Do not use the Warmer system if the system failure alarm is activated. Remove
the unit from service and refer to qualified personnel for repair.

w

w

w

w

w

Cautions

Only competent individuals trained in the repair of this equipment should
attempt to service it as detailed in the service manual. The Service Manual
provides detailed information solely for use by individuals having proper
knowledge, tools and test equipment, and for service representatives trained
by Ohmeda Medical.

General Precautions

6600−0371−000−DWG

100

MANUAL−DOC, User Reference Manual, OM Panda, English−default

Reproduced from the electronic master in MATRIX

Summary of Contents for Panda

Page 1: ...mer OperationandMaintenanceManual 6600 0371 000 DWG 100 MANUAL DOC User Reference Manual OM Panda English default Reproduced from the electronic master in MATRIX Approved 2009 6 4 16 13 EET DST Barbar...

Page 2: ...to the nearest Ohmeda Medical Regional Service Center This Product or any of its parts should not be repaired other than in accordance with written instructions provided by Ohmeda Medical and by Ohme...

Page 3: ...CheckoutProcedure 2 4 A Control Unit Check 2 4 B ElapsedTimerCheck 2 5 C ObservationLightCheck 2 5 D Raise and Lower Bed Switch Check 2 5 E Interlock Switch Check 2 6 F Power Failure Memory and Batter...

Page 4: ...y 5 5 5 5 Battery Test 5 5 5 6 Maintenance Schedule 5 6 A Operator Maintenance 5 6 B Service Maintenance 5 6 6 Illustrated Parts Appendix Temperature Conversion Chart A 1 Servo Mode Algorithm A 2 Infa...

Page 5: ...is used when the possibility of injury to the patient or the operator exists Symbols Attention Consult the manual for more information Type B equipment FunctionalEarthTerminal ProtectionEarthTerminal...

Page 6: ...iv 6600 0371 000 10 27 03 Table of Contents iv Notes 6600 0371 000 DWG 100 MANUAL DOC User Reference Manual OM Panda English default Reproduced from the electronic master in MATRIX...

Page 7: ...rocedures Mechanical and Control Unit while a patient occupies the Warmer Complete the Checkout Procedures section of this manual before putting the unit into operation If the warmer fails any portion...

Page 8: ...vi 6600 0371 000 10 27 03 Table of Contents vi Notes 6600 0371 000 DWG 100 MANUAL DOC User Reference Manual OM Panda English default Reproduced from the electronic master in MATRIX...

Page 9: ...accessory items 1 3 Heater Assembly Thewarmerisdesignedtowarminfantsonabedsurfacelocatedbeneaththeheaterassembly WARNING w Bed to heater spacing less than the recommended 64 cm will result in incor r...

Page 10: ...le bed platformallowsTrendelenburgandreverseTrendelenburgpositioning Thehydraulic system for the tiltable bed provides a smooth dampened motion to avoid disturbing the patient 1 6 Bassinet Two bassine...

Page 11: ...rocedure WARNING w Before using the Ohmeda Panda Baby Warmer read this entire manual Attempting to use this device without a thorough understanding of its operation may result in patient or user injur...

Page 12: ...back to the normal position Check for a smooth rotation C Mechanical Checks 1 Check the operation of the bed sides The bed sides should operate smoothly WARNING w Regularly inspect the bed side panel...

Page 13: ...ator Alarm Lights Alarm Silence Mode Switch Mode Indicators Bed Controls Bed controls present on elevating bed models only CI 14 096 Figure 2 1 Wall mount pin insertion 1 Verify that the warmer is rig...

Page 14: ...prompt tones sound and the power display flashes 2 Adjust the heat output with the increase M and decrease touch switches to attain the high and low limits as indicated by the power display 3 Connect...

Page 15: ...ch switch to activate the elapsed timer Verify that the timer starts operating 2 Press the Apgar tones On Off touch switch for the Apgar tones Verify that the indicator light for the Apgar tones is ex...

Page 16: ...10 minutes test or partially charged to pass the two minute test If the battery is defective replace it Refer to the service manual There is no maintenance required for the battery The battery has a...

Page 17: ...erature sensed at the skin tempera ture probe tip The skin temperature probe must be properly connected to the warmer and the infant for an accurate patient temperature measurement The range of temper...

Page 18: ...st take the responsibility for making heater adjustments or environmental changes in response to the following conditions patient size and weight bed to heater head height differences changes in the e...

Page 19: ...alarm tone is sounded at one minute and every five minute interval after the elapsed timer is started LightOn Off This switch activates the observation light located in the heater assembly WARNING w...

Page 20: ...owerFailureAlarm Probe Failure Alarm The probe failure alarm is only active in the servo mode of operation The alarm activates when the skin temperature probe fails electrically due to an open or shor...

Reviews: