42
Conducting a measurement
After all the equipment has been prepared and the GeoSeeker Mini has been powered on, you can start
to process the measurement according to the instructions of chapter 5.1 “New Measurement” on page
Under certain conditions the control unit might display some warnings or error messages during the
measurement:
•
Power induction failed
If this screen appears the device can not inject
power into the underground. Please check all
connections and make sure there are no cable breaks
or unconnected plugs.
Then try again to measure the current scan point.
•
Voltage failure
If this screen appears the device cannot measure any
voltage between the electrodes. Please check all
connections and make sure there are no cable breaks
or unconnected plugs.
Then try again to measure the current scan point.
•
General measurement error
Under extreme conditions it might be possible that
your device is not able to induct enough power into
the underground. After several trials the device will
abort the current scan and will show this screen.
You should replace the power electrodes for at least
1 - 2 meters and try again. Alternatively you can try
to water the electrodes to get better electrical
contact to the surrounding soil as explained in