20
Figure 22. How to reach the Sensor Calibration Submenu
9.2.1.2.1
Offset (with exthernal meter)
Press on Offset icon
to calibrate the instrument by using an external CO
2
meter (NOT hereby
provided).
Figure 23. Sensor Calibration Submenu
To run the gas sensor calibration through an external gas meter follow the indications below:
1.
Disconnect the blue tubing inserted into the
“Gas Output”
port. Remove it by keeping the black
ring of this connector pushed in; you should feel no resistance in pulling this tubing out. If you
do make sure you are properly pushing in the black ring.
Output Gas
External Gas Meter
Tubing not supplied
CO2-UNIT-BL
Figure 24. External Gas Meter Connection