26
9.2.1.3
Flowrates
Press the Flowrates icon to check how many liters per minute of Air and CO
2
the system is consuming
(see Figure 37 a). Additionally, this panel allows you to select the flowrate of the gas mixture you want to send to
your incubator/device. Press the
“Total” flowmeter
to set the gas flowrate in the range 0.0÷1.0 L/min using the
slider, as shown in Figure 38.
Figure 37. Flowrates View (a
–
b).
1
2
Figure 38. How to change the total flow rate.
When setting the flowrate at 0.0 l/min the gas flowrate will be paused and output flow will be
interrupted. In this working mode three dots will appear in Homepage instead of CO
2
% current
value and the indicator will turn yellow.
When using the OKO-AIR-PUMP-BL as Air Source, the maximum flowrate value you can set is 0.4
l/min.
a
b