29
Figure 44. Warning shown if you try to change set point during a gas cycle
9.2.1.5
Status
Press the Status icon
to check the CO2-UNIT-BL performance (see
Figure 45 a). Most of the parameters displayed are technical/control parameters that you may be asked to share in case you
have requested for Okolab support (see
a
b
Figure 45. Status page (a
–
b).
The values
that may be of interest to users are “Flow” and “Gas %”. All other parameters are just to be
communicated to our technicians in case you may need Okolab support.
9.2.1.6
Digital Input
Digital Input feature is available only for TTL ready (optional) controller.
TTL feature allows to change setpoint between two specified values using a TTL signal (see Figure 46 b).
The TTL is a digital input for our Units and has only two values: High and Low. The setpoint values for High TTL
value and for Low TTL value can be set using OKO-TOUCH, via GPIO connector.
When TTL Control is enabled, the user cannot change the setpoint using the dialog in the main page. In
this case, the setpoint value is read by the external TTL signal.