8
3.
MINI-USB-OTG
cable. To connect a USB pen to the OKO-TOUCH for data download.
−
CO
2
tank with safety valve/pressure gauge regulator to accept rigid tubing 6 mm OD. Gas source must be
Standard Purity CO2 (coded as 4.5 that means 99.995 % of CO2) and humidity-free gas.
−
Air tank or compressed air line with safety valve/pressure gauge regulator (degree of filtration 20 µm and
condensate drain) to accept rigid tubing 6 mm OD.
Note
►
An alternative solution for Air supply is the OKO-AIR-PUMP-BL.
6
Optional Equipment
It can be equipped with:
•
OKO-AIR-PUMP-BL
(optional)
that includes
:
▪
TUBE B
. 2 m long tube composed of 6 mm OD rigid tube (0.5 m long) connected to 4 mm ID silicon
transparent tube (0.5 m long) then connected to an hydrophobic filter 0.2 micron porous, connected
to 4mm ID silicon transparent tube 0.5 m long and then connected to 6 mm OD rigid tube (0.5 m
long). Use the TUBE B to connect the OKO-AIR-PUMP-BL to the port
labeled “Air In” on the rear panel
of the CO2 Controller.
OKO-AIR-PUMP-BL
TUBE B
Figure 2. CO2 UNIT BL with OKO AIR PUMP BL
•
HM-ACTIVE
(optional)
that includes:
▪
Humidity Sensor Lid
with Temperature control.
▪
Humidity bottle
and its Humidifier
(
Heated Base
)
▪
Heated and Insulated Tube