After that, you can select one of the saved files using the
and
buttons. After the desired file is selected, you need to press the
button to view
or delete this file. In this case, the window will change and will be as shown below:
In this window, using
,
and
buttons you need to select the
action that you want to perform with the selected file. The file can be opened or deleted.
Additionally, in this menu it is possible to delete all saved files.
To exit the menu, press the
button the required number of times.
Presets and calibration
There are a number of steps to take before starting measurements:
1)
In the main window of the device, select the material of the test sample from
the list (see “The main window” section).
2)
Select the temperature of the test sample (see “The main window” section).
The speed of sound in a material is temperature dependent. Therefore, it is
necessary to try to set the temperature of the test sample as accurately as
possible so that the device takes into account the change in the speed of
sound. This is especially important for measuring the thickness of heated
samples.
ATTENTION!
When measuring the thickness of a sample with a surface temperature of more
than 80 °C, it is necessary to use the EMT40004T converter.