EMA Thickness gauge EM4000. User Manual
+7(812) 385-54-28
info@oktanta-ndt.ru
22
and press the
button.
The third step is thickness calibration.
To do this, select the appropriate menu item and press the
button.
After that, a field appears on the device screen, in which you can enter the
value of the thickness of the reference coating. The very value of this
thickness is indicated on the sample supplied. Install the instrument on the
sample as shown in the figure below:
Positioning the device for zero calibration
and press the
button.
After completing these three points, the device should correctly measure the
thickness of the plastic coating on the sample from the delivery set.
ATTENTION!
Calibrate the coating thickness measurement channel only in the
sequence specified in this section and only with the EMT40001
converter installed.