PEC Flaw Detector PE4332 Operation Manual
+7(812) 385-54-28
info@oktanta-ndt.ru
15
Scanning stop
To stop scanning, select "Stop scanning" menu item and press on the
key on
the Device keyboard.
Scanning continuation
To continue scanning, select "Continue scanning" menu item and press on the
key of the Device keyboard. This function can be used not only to continue the
stopped scanning, however, to continue the previous scans that were not completed.
For example, the user may have scanned only part of the test item and then he/she can
return to the test item few days later and continue scanning. To do this, the user shall
open the archive with non-completed scan (see Section "Main menu" / "Open file"),
open menu of scanning start/stop/continuation and select "Continue scanning". Do not
forget that in this case calibration shall be performed prior to scanning continuation.
WARNING! You can get access to "Scanning" menu through the
Device main menu
key.
Calibration
Figure 15 shows the Device calibration window. To perform calibration, install
the sensor on the test item section with thickness known, set this thickness in the
Device and press on the
key on the Device.