4.5. Replacing the fuse 

 

 

It is mandatory that spare parts must be guaranteed original OLDHAM parts as, 
otherwise, the reliability of the equipment could be adversely affected. 

 

 

 

The fuse (Fig. 9, item 1) must be replaced by qualified personnel only. 

The fuse is, and must be, in compliance with CEI 127, with time-delay, low breaking capacity  

and a voltage of 250 V ~. 

- Fuse 5x20- T125 mA 250 V- OLDHAM reference = 6154701 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

FIGURE 9 

Item 1 

Summary of Contents for SV 4B

Page 1: ...Surveyor 4B Part Number NPS4BGB Version C The Fixed Gas Detection People User Manual...

Page 2: ......

Page 3: ...HEIR SPECIFIC USE C C L L A A U U S S E E S S C C O O N N C C E E R R N N I I N N G G P P R R O O P P E E R R T T Y Y Drawings plans specifications and information included in this document contain co...

Page 4: ...ly as concerns electricity and fitting connections and connection to the power network Any modification of the equipment or the use of parts not specified as original manufacturer s parts could lead t...

Page 5: ...tallation 11 IV OPERATING INSTRUCTIONS 12 4 1 Switching on 13 4 2 Switching off 13 4 3 Alarms 13 4 3 1 GAS alarm 13 4 3 2 FAULT alarms 14 4 4 Adjustments 14 4 4 1 Adjusting the gas alarm thresholds 15...

Page 6: ...6...

Page 7: ...ests on the top item 1 indicator lights on the FRONT face item 2 connections at the bottom item 3 II MOUNTING OF THE SURVEYOR 4B The surveyor 4B appliance fitted on its symmetrical DIN rail can be ins...

Page 8: ...8...

Page 9: ...r 12 VCC 2 3 1 Alternative power supply 230 volt AC power supply The Surveyor 4B must be protected on the upstream side by a two pole earth leakage circuit breaker 1A The response curve must be of typ...

Page 10: ...to the first gas alarm threshold and to the FAULT alarm This relay is equipped with SPDT contacts available on the SV 4B terminal block item 3 Fig 5 Relay 2 REL 2 which is negative safety mode corres...

Page 11: ...di is sp pl la ay y l li ig gn nt t Gas valve closed Breaking capacity of rating relays switch contact 120VA ou 30W Important The relay s contacts are shown on the SV 4B label unit switch off The AL1...

Page 12: ...a al lv ve e s st ta at tu us s d di is sp pl la ay y l li ig gn nt t Gas valve closed Breaking capacity of rating relays switch contact 120VA ou 30W Important The relay s contacts are shown on the SV...

Page 13: ...sitive safety position 4 2 Switching off The SV 4B is switched off by cutting off the power supply on an electrical equipment cabinet 4 3 Alarms 4 3 1 GAS alarm The SV 4B has two adjustable GAS alarm...

Page 14: ...for each sensor for the purpose of verifying sensor and alarm operation and is not intended to be a measure of the accuracy of the instrument OLDHAM further recommends that a full instrument calibrat...

Page 15: ...inimum of 25 LEL should be injected Adjust the alarm potentiometer item 1 Fig 7 to trigger the first threshold AL1 the red LED AL1 flashes for 7 seconds and then remains on in steady mode trip the cor...

Page 16: ...ou read 1 000 mV for instance If you set the alarm potentiometer to read 1 500 mV the alarm is triggered at 1 5 methane Signal mV mV mVx 1500 1000 1 1000 Or Threshold 1 1500 1000 15 x mV mV 4 4 2 Ackn...

Page 17: ...ure8 4 4 4 Checking the sensitivity Necessary when a cell is replaced At least twice a year Prepare the calibration kit and secure the gas input pipe to the detector Adjust the flow rate of the refere...

Page 18: ...is satisfactory Cell sensitivity is inadequate Can AL 1 be triggered by adjusting the alarm potentiometer item 2 Fig 8 Cell sensitivity is now satisfactory Cell sensitivity is inadequate replace the...

Page 19: ...ise the reliability of the equipment could be adversely affected The fuse Fig 9 item 1 must be replaced by qualified personnel only The fuse is and must be in compliance with CEI 127 with time delay l...

Page 20: ...20...

Page 21: ...ms Mounting On symmetrical DIN rail Miscellaneous Technology Mains visual indicator Housing Warranty SMC surface mounted component Green LED NORYL 1 year Dimensions 58 x 105 x 90 mm Weight 0 360 kg Ca...