29
CKX41/CKX31
Fig. 34
Fig. 35
9
Mounting the Filters
(Figs. 34 & 35)
Let the filters cool down sufficiently before replacing them.
Take out the filter holder
1
and insert the required filters
2
.
# Push the filter down to the bottom as shown in Fig. 35 so that it does
not tilt. If the filter is inclined or is not pushed down to the bottom, it
may fall off the filter holder.
}More than one filter can be stacked in the filter holder. You can mount as
many as you like, as long as the total thickness does not exceed 11 mm.
@
²
Fig. 36
Fig. 37
10
Connecting the Cables and Power Cord
(Figs. 36 & 37)
Cables and cords are vulnerable when bent or twisted. Never subject
them to excessive force.
Make sure that the main switch of the power supply is set to “ ”
(OFF) before connecting cables.
1. Connect the plug
1
of the illumination column firmly to connector
2
on
the rear of the microscope.
Do not connect anything other than the plug
1
of the cord from the
illumination column to connector
2
. Doing so will cause equipment
failure.
Always use the AC power adapter and power cord provided by
Olympus. If no power cord is provided, please select the proper
power cord by referring to the section “PROPER SELECTION OF THE
POWER SUPPLY CORD” at the end of this instruction manual.
2. Connect the power cord connector
3
to connector
4
firmly. (Fig. 36)
3. Connect the power cord plug
5
to a wall outlet
6
. (Fig. 37)
Be sure to supply power from a grounded, 3-conductor power outlet
using the proper power cord. If the power outlet is not grounded
properly, Olympus can no longer warrant the electrical safety perfor-
mance of the equipment.
If the power cord or a connection cable comes in contact with the
lamp socket or surrounding equipment, the cord or cable may melt
and result in shock hazard. To prevent this, distribute the cords and
cables apart from the lamp socket.
@
²
³
|
5 6
Correct
Incorrect